Linear accuracy of cone-beam computed tomography and a 3-dimensional facial scanning system: An anthropomorphic
Song Hee Oh1, Ju Hee Kang1, Yu-Kyeong Seo1
1Department of Oral and Maxillofacial Radiology, Graduate School, Kyung Hee University, Seoul, Korea.
Purpose:
This study was conducted to evaluate the accuracy of linear measurements of 3-dimensional (3D) images generated by cone-beam computed tomography (CBCT) and facial scanning systems, and to assess the effect of scanning parameters, such as CBCT exposure settings, on image quality.
Materials And Methods:
CBCT and facial scanning images of an anthropomorphic phantom showing 13 soft-tissue anatomical landmarks were used in the study. The distances between the anatomical landmarks on the phantom were measured to obtain a reference for evaluating the accuracy of the 3D facial soft-tissue images. The distances between the 3D image landmarks were measured using a 3D distance measurement tool. The effect of scanning parameters on CBCT image quality was evaluated by visually comparing images acquired under different exposure conditions, but at a constant threshold.
Results:
Comparison of the repeated direct phantom and image-based measurements revealed good reproducibility. There were no significant differences between the direct phantom and image-based measurements of the CBCT surface volume-rendered images. Five of the 15 measurements of the 3D facial scans were found to be significantly different from their corresponding direct phantom measurements (P<.05). The quality of the CBCT surface volume-rendered images acquired at a constant threshold varied across different exposure conditions.
Conclusion:
These results proved that existing 3D imaging techniques were satisfactorily accurate for clinical applications, and that optimizing the variables that affected image quality, such as the exposure parameters, was critical for image acquisition.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Voltammetric Techniques: Linear-Scan (E vs Time)
Improving Translational Accuracy
Leaky Scanning
Uncertainty in Measurement: Accuracy and Precision