RESOLUTION, DEPTH OF FIELD, AND PHYSICIAN SATISFACTION DURING DIGITALLY ASSISTED VITREORETINAL SURGERY
William R Freeman1,2, Kevin C Chen1,2, Joseph Ho1,2
1Department of Ophthalmology, Shiley Eye Institute, University of California San Diego, La Jolla, California.
Purpose:
To evaluate depth of field, lateral resolution, and image quality of a heads-up 3D visualization system for vitreoretinal surgery using physician survey and optical measurement outcomes.
Methods:
Depth of field and lateral resolution were compared between the standard ocular viewing system and the digital 3D system at ×5, ×13, and ×18 magnification by 6 retinal surgeons. Optical techniques were used as well as a survey of surgeon impression. Surgeon impression surveys were performed after 6 weeks of surgical use of the device.
Results:
Physician questionnaire survey scores for depth of field at high magnification were better for the digital 3D system and equivalent for all other categories. Measured lateral resolution was 36.7 mm and 16.6 mm at ×5 magnification (P < 0.001), 14.3 mm and 6.4 mm at ×13 magnification (P < 0.001), and 9.8 mm and 4.2 mm (P < 0.001) at ×18 magnification for the digital 3D and oculars, respectively. Measured depth of field was 4.00 mm and 6.78 mm at ×5 magnification (P = 0.027), 0.72 mm and 0.86 mm at ×13 (P = 0.311), and 0.28 mm and 0.40 mm at ×18 magnification (P = 0.235) for the oculars and digital 3D, respectively.
Conclusion:
Lateral resolution of the digital 3D system was half that of the ocular viewing system and there was some improvement in depth of field with the digital system. Surgeon impression suggested that the digital system was superior when evaluating depth of field at high magnification.
Related Concept Videos
Uniform Depth Channel Flow
Depth Perception and Spatial Vision
Uniform Depth Channel Flow: Problem Solving
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Field Effect Transistor
Chromatographic Resolution
The effectiveness of separation can be evaluated by determining the level of separation between two neighboring peaks in a chromatogram, which represents the individual components of a sample.
In chromatography,...