Novel algorithm to screen for heart murmurs using computer-aided auscultation in neonates: a prospective single
Renata Grgic-Mustafic1,2, Nariae Baik-Schneditz1,2, Bernhard Schwaberger1,2
1Research Unit for Neonatal Micro- and Macrocirculation, Department of Pediatrics and Adolescent Medicine, Medical University of Graz, Graz, Austria.
Background:
Automated detection of heart murmurs with computer-aided auscultation is not yet in clinical routine use. Aim of this study was to test sensitivity and specificity of a novel prototype algorithm in automated detection of heart murmurs from digitally recorded phonocardiograms in neonates admitted at the Neonatal Intensive Care Unit.
Methods:
In a prospective pilot observational study from November 2012 to December 2013 auscultations by pediatricians and computer aided auscultation were performed within 12 hours of neonatal echocardiography. Echocardiography was defined as pathological when resulting in any clinical consequences or causing murmur. Phonocardiograms and auscultation were defined as pathological if a murmur was detected. Phonocardiograms were analyzed offline with a novel algorithm prototype (CSD Labs, Graz, Austria) for detection of murmurs in neonates in a first run and with an optimized algorithm in a second run and were compared with echocardiography. Sensitivity and specificity of auscultation by pediatrician and computer aided auscultation were analyzed.
Results:
Thirty-six neonates (gestational age: 36±3 weeks) were included. Twenty-three (64%) neonates had pathological or murmur causing findings in echocardiography (positive echocardiography). Sensitivity and specificity of auscultation by pediatrician were 17% and 100%, respectively. In comparison to auscultation by pediatrician sensitivity of first run and second run were significantly higher with 70% and 83%, respectively. Specificity of first run and second run were 77% and 85%, respectively.
Conclusions:
Phonocardiogram analysis using the novel algorithm prototype had a higher sensitivity than auscultation by pediatrician in detecting positive echocardiography findings in neonates.
Related Concept Videos
Observational Studies
There are three types of observational studies – Prospective, retrospective, and cross-sectional.
Prospective Study
Prospective studies, also known as longitudinal or cohort studies, are carried out by collecting future data from groups sharing similar characteristics. One...
Pilot and Numeric Relaying
Lattice Centering and Coordination Number
Types of Unit Cells
Imagine taking a large number of identical...
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Naturalistic Observations
Trial and Error and Algorithm