Measurement of Urinary Bladder Pressure: A Comparison of Methods
Ingelin Clausen1, Lars Geir W Tvedt2, Thomas Glott3
1Department of Microsystems and Nanotechnology, SINTEF Digital, NO-0314 Oslo, Norway. ingelin.clausen@sintef.no.
Abstract:
Pressure is an essential parameter for the normal function of almost all organs in the human body. Measurement of pressure is therefore highly important in clinical practice and medical research. In clinical practice, pressures are often measured indirectly through a fluid line where the pressure is transmitted from the organ of interest to a remote, externally localized transducer. This method has several limitations and is prone to artefacts from movements. Results from an in vitro bench study comparing the characteristics of two different sensor systems for bladder assessment are presented; a new cystometry system using a MEMS-based in-target organ sensor was compared with a conventional system using water-filled lines connected to external transducers. Robustness to measurement errors due to patient movement was investigated through response to forced vibrations. While the new cystometry system detected real changes in applied pressure for excitation frequencies ranging from 5 Hz to 25 Hz, such small and high-frequency stimuli were not transmitted through the water-filled line connected to the external transducer. The new sensor system worked well after a resilient test at frequencies up to 70 Hz. The in-target organ sensor system will offer new possibilities for long-term monitoring of in vivo pressure in general. This opens up the possibility for future personalized medical treatment and renders possible new health services and, thereby, an increased patient empowerment and quality of life.
Related Concept Videos
Urinary Bladder
In males, the bladder is situated in front of the rectum, while in females, it is positioned anterior to the vagina and uterus. The bladder floor contains an inverted triangular area called the trigone, defined by the two ureteric...
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...