Background And Objectives:
During storage, red blood cells (RBCs) undergo physicochemical changes which affect the quality, function, and in vivo survival of transfused packed RBCs (pRBC). Changes include decreased 2,3-diphosphoglycerate (2,3-DPG) levels, decreased ATP, changes in mechanical properties and oxidative injury. RBC rejuvenation is a method used to increase levels of 2,3-DPG and ATP in pRBCs. This process requires incubating the pRBCs with a rejuvenation solution and subsequent washing. Standard blood bank protocols using the COBE 2991 Cell Processor require several hours of preparation. The objective of this study was to verify if a bedside protocol for rejuvenating pRBC and washing with the Sorin Xtra autologous cell salvage system could be used.
Materials And Methods:
Outdated pRBC units were obtained and rejuvenated in a model operating suite using a dry air incubator for 1 h at 37°C. Six units of pRBCs were pre-diluted with saline (1000 ml) and six units were not pre-diluted with saline. All units were washed with normal saline (1000 ml) using an apheresis-design cell salvage device in manual mode and wash volume set to 3000 ml. Samples were collected and analyzed for standard RBC quality parameters at baseline and post-wash.
Results:
Total pRBC wash efficiency was 94% ± 12% at a final hematocrit of 67.7 ± 5.9% while maintaining post-wash hemolysis 0.24 ± 0.12 %. Pre-dilution prior to washing did not confer statistically significant differences in final RBC quality parameters with the notable exceptions of calculated hemolysis and supernatant potassium levels (P < 0.05). The washing process can be completed within 10 min. The post-wash RBC parameters are appropriate for immediate transfusion to patients.
More Related Videos
11:49Autologous Endothelial Progenitor Cell-Seeding Technology and Biocompatibility Testing For Cardiovascular Devices in Large Animal Model
Published on: September 9, 2011
09:49A Simple Red Blood Cell Lysis Method for the Establishment of B Lymphoblastoid Cell Lines
Published on: January 14, 2017
Related Concept Videos
Protein Buffers in Blood Plasma and Cells
Certain amino acids can exist in a zwitterion state at a...
Red Algae
Adult Stem Cells
What are Cells?
Basic Characteristics of Cells
A living cell has a plasma membrane, a bilayer of lipids that separates the aqueous solution inside the cell called the cytoplasm from the outside environment.
Furthermore, a living cell possesses genetic information...
Voltaic/Galvanic Cells
Spontaneous redox reactions occur abundantly in nature. The chemical reaction occurring in a disposable AA battery powering our remote controls is one such example of a spontaneous redox reaction. Another example is the immersion of coiled copper wire into an aqueous silver nitrate solution. The reaction shows a gradual, visually impressive color change from colorless to bright blue and the formation of a grey precipitate on the copper wire. In this experiment,...
Non-ohmic Devices
Consider a simple circuit consisting of a battery, a diode, and a resistor. A...