Impact of new ICRU Report 90 recommendations on calculated correction factors for reference dosimetry
Damian Czarnecki1, Björn Poppe, Klemens Zink
1Institute of Medical Physics and Radiation Protection, University of Applied Sciences Giessen, Giessen, Germany. University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl von Ossietzky University, Oldenburg, Germany.
Abstract:
In 2016 the ICRU published a new report dealing with key data for ionizing radiation dosimetry (ICRU Report 90). New recommendations have been made for the mean excitation energies I for air, graphite and liquid water as well as for the graphite density to use when evaluating the density effect. In addition, the ICRU Report 90 discusses renormalized photoelectric cross sections, but refuses to give a recommendation on the use of renormalization factors. However, the Consultative Committee for Ionizing Radiation recommends to use renormalized photoeffect cross sections. Goal of the present work is to evaluate the impact of these new recommendations on clinical reference dosimetry for high energy photon and electron beams. The beam quality correction factor k Q was calculated by Monte Carlo simulations for compact and parallel plate ionization chambers. In case of photons seven phase space files from clinical accelerators and twelve spectra taken from literature from 4 MV to 24 MV and additionally a 60Co source were applied. As electron source thirteen electron spectra available in literature were used in the range of 4 MeV-21 MeV. The new ICRU recommendations have a small impact on Monte Carlo calculated k Q values for the chosen ionization chambers in the range of 0.1%-0.35% only-the difference increases for higher photon energies. The impact of the ICRU Report 90 recommendations on Monte Carlo calculated stopping power ratios s w,a , perturbation factors p and beam quality correction factors k Q was investigated and confirmed a decrese of s w,a by a fraction of a percent for photon and electron beams. This study indicates that the impact of the new ICRU recommendation is within 0.35%. The determined deviations should be taken into account, when widely published Monte Carlo calculated values are examined.
More Related Videos
07:31Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Power Factor Correction
Distance Corrections
Calculating Standard Free Energy Changes
Data Reporting and Recording
Inertial Frames of Reference
Non-inertial Frames of Reference