Biomechanical Comparison of Screw Fixation Versus a Cortical Button and Self-tensioning Suture for the Latarjet
Matthew T Provencher1,2, Zachary S Aman1, Christopher M LaPrade1
1Steadman Philippon Research Institute, Vail, Colorado, USA.
Background:
Metal screws are traditionally used to fix the coracoid process to the glenoid. Despite stable fixation, metal screws have been associated with hardware complications. Therefore, some studies have advocated for suture button fixation during the Latarjet procedure to reduce the complications associated with screw fixation.
Purpose:
To biomechanically evaluate the ultimate failure load of a cortical button and self-tensioning suture versus metal screws for coracoid graft fixation during the Latarjet procedure.
Study Design:
Controlled laboratory study.
Methods:
Eight matched pairs of fresh-frozen, male cadaveric shoulders (N = 16) underwent the Latarjet procedure. The shoulders of each pair were randomly assigned to 1 of 2 groups: fixation using two 3.75-mm cannulated, fully threaded metal screws or fixation using a double suture button construct. Specimens were secured in a dynamic testing machine and cyclically preconditioned from 2 to 10 N at 0.1 Hz for 10 cycles. After preconditioning, specimens were pulled to failure at a normalized displacement rate of 400% of the measured gauge length per minute. The ultimate failure load and mechanism of failure were recorded for each specimen.
Results:
The mean ultimate load to failure for screw fixation (226 ± 114 N; 95% CI, 147-305 N) was not significantly different from that for suture button fixation (266 ± 73 N; 95% CI, 216-317 N) (P = .257). The mean strain at failure for screw fixation (63% ± 21%; 95% CI, 48%-77%) was not significantly different from that for suture button fixation (86% ± 26%; 95% CI, 69%-104%) (P = .060). The most common mechanism of failure for the screw fixation method was at the bone block drill holes, while an intramuscular rupture at the clamp-muscle interface occurred for the suture button construct.
Conclusion:
The screw and suture button fixation techniques exhibited comparable biomechanical strength for coracoid bone block fixation of the Latarjet procedure.
Clinical Relevance:
Metal screws have been reported to be a large contributor to intraoperative and postoperative complications. Therefore, given the results of the current study, a suture button construct may be an alternative to metal screw fixation during the Latarjet procedure. However, further clinical studies are warranted.
Related Concept Videos
Tension
Screw: Problem Solving
To evaluate the...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Surface Tension, Capillary Action, and Viscosity
The various IMFs between identical molecules of a substance are examples of cohesive forces. The molecules within a liquid are surrounded by other molecules and are attracted equally in all directions by the cohesive forces within the liquid. However, the molecules on the surface of a liquid are attracted only by about one-half as many molecules. Because of the unbalanced molecular attractions on the surface molecules, liquids contract to form a shape that minimizes the number...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...