Impact of patient-specific morphologies on sinus flow stasis in transcatheter aortic valve replacement: An in vitro
Hoda Hatoum1, Jennifer Dollery2, Scott M Lilly3
1Department of Biomedical Engineering, The Ohio State University, Columbus, Ohio.
Objective:
The goal of this study is to evaluate how sinus flow patterns after transcatheter aortic valve replacement in realistic representative patient roots vary. Sinus flow can affect transcatheter aortic valve operation and likely leaflet thrombosis occurrence due to stasis and poor washout. How the interaction between transcatheter aortic valve and representative patient aortic roots affects sinus hemodynamics is important to establish for future individualization of transcatheter aortic valve replacement therapy.
Methods:
Two representative patient aortic roots were selected, segmented and 3-dimensional printed followed by deployment of Medtronic CoreValve (Medtronic Inc, Irvine, Calif) and Edwards SAPIEN (Edwards Lifesciences, Irvine Calif) transcatheter aortic valves. Sinus hemodynamics were assessed in vitro using high spatio-temporal resolution particle-image-velocimetry. Detailed sinus vortex tracking, shear stress probability density functions, and sinus washout were evaluated and assessed as a function of valve type and representative patient morphology as independent case studies.
Results:
Peak velocity in the sinus with SAPIEN valve was approximately 3 times higher than with CoreValve for both models (0.30 ± 0.02 m/s and 0.34 ± 0.041 m/s vs 0.13 ± 0.01 m/s and 0.10 ± 0.02 m/s) (P < .01). Between representative patient models, vorticity magnitudes were significantly different (75 ± 1.1 s-1, 77 ± 3.2 s-1, 109 ± 2.3 s-1, and 250 ± 4.1 s-1) (P < .01) regardless of valve type. Sinus blood washout characteristic as a function of cardiac cycles was strongly both patient related and valve specific. Fluid dynamics favored shear stresses and washout characteristics due to a smaller sinus and sinotubular junction, further amplified by the SAPIEN valve.
Conclusions:
Sinus flow dynamics are highly sensitive to aortic root characteristics and transcatheter aortic valve aortic root interaction. Differences in sinus-flow washout and stasis regions between representative patient models may be reflected in different risks of leaflet thrombosis or valve degeneration.
More Related Videos
Related Concept Videos
Heat Flow and Specific Heat
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Impact of Groups on Groups
ECG Interpretation of Arrhythmias I: Sinus Arrhythmias
Types of Arrhythmias
Sinus Node Arrhythmias
Sinus Bradycardia: Originating from the sinoatrial (SA) node, sinus bradycardia involves slower impulses, resulting in a heart rate of less than 60 beats per minute (bpm). Causes include sleep, vagal stimulation, beta-blockers, hypothyroidism,...
Impact
When particles with different initial velocities collide, they induce deformation by applying equal and opposite impulses. At the point of maximum deformation, the particles move together with...
Continuous Renal Replacement Therapy