Related Experiment Video
Updated: Feb 8, 2026

Characteristics of Precipitation-formed Polyethylene Glycol Microgels Are Controlled by Molecular Weight of Reactants
Published on: December 23, 2013
Less Midterm Damage and Oxidation Are Seen in Retrieved Highly Crosslinked Ultrahigh-Molecular-Weight Polyethylene
Alexandra Stavrakis1, Lydia Weitzler1, Timothy Wright1
1Hospital for Special Surgery, 535 E. 70th St., New York, NY 10021 USA.
Background:
Highly crosslinked ultrahigh-molecular-weight polyethylene (XLPE) shows reduced wear in total hip arthroplasty compared to direct compression-molded polyethylene (compPE); however, minimal research evaluating polyethylene damage in XLPE tibial inserts in total knee arthroplasty exists.
Questions/Purposes:
We evaluated damage and material properties in retrieved XLPE components at midterm (≥ 2.5 years) follow-up.
Methods:
We identified 19 XLPE tibial inserts with ≥ 30 months in vivo using our institutional review board-approved implant retrieval system; 19 compPE retrieved inserts were matched based on age at index surgery, body mass index, sex, and length of implantation. Articular surface damage was assessed using a subjective grading system. Swell ratio testing and Fourier-transform infrared spectroscopy were used to measure crosslink density (XLPE) and oxidation (XLPE, compPE), respectively, at loaded and unloaded surface and subsurface regions.
Results:
CompPE inserts had higher overall damage than XLPE inserts, specifically at the post of posterior-stabilized inserts. The XLPE inserts had lower crosslink density at the loaded surface (0.159 mol/dm3) than either unloaded region (0.183 mol/dm3). CompPE peak oxidation index (OI) was greater than XLPE peak OI in the loaded and unloaded surface regions (1.67 vs. 0.61 and 1.38 vs. 0.46, respectively).
Conclusions:
Surface damage and oxidation are reduced in XLPE inserts compared to compPE at midterm follow-up. Peak OI greater than 1.0 in the compPE group suggests that mechanical-property degradation had occurred, a likely cause for increased damage. Longer-term retrievals will determine whether these trends continue. Based on midterm results, XLPE shows an advantage over compression molded PE in total knee arthroplasty.
Related Concept Videos
Porin Insertion in the Outer Mitochondrial Membrane
Three models describe the assembly of porins by the SAM complex and their insertion into the outer membrane. Model 1 suggests that porins are assembled outside the SAM channel as the...
Polymers: Defining Molecular Weight
The number average molecular weight (Mn) is the summation of the number...
Polymers: Molecular Weight Distribution
Insertion of Single-pass Transmembrane Proteins in the RER
Integral transmembrane proteins possess transmembrane and extra membrane domains. The transmembrane domains are primarily made of 20-25 hydrophobic amino acids arranged in a helical secondary confirmation. These...
Insertion of Multi-pass Transmembrane Proteins in the RER
The multipass transmembrane proteins are the type IV integral membrane proteins with multiple topogenic sequences determining their spatial arrangement in the ER membrane. Nearly all multipass proteins lack a cleavable signal sequence and use...
Density, Specific Weight, Specific Gravity and Compressibility of Fluid
Specific weight represents the weight per unit volume and is calculated by multiplying...