Agreement and repeatability of central corneal thickness measurements by four different optical devices and an
Nesrin Buyuktortop Gokcinar1, Erhan Yumusak2, Nurgul Ornek2
1Department of Ophthalmology, Faculty of Medicine, Kırıkkale University Hospital, Yahşihan, Kırıkkale, Turkey. tortopn@kku.edu.tr.
Purpose:
To compare the repeatability and agreement of central corneal thickness (CCT) measurements by spectral-domain optical coherence tomography (OCT), corneal topography (CT) with a combined Scheimpflug-Placido system, optical biometry (OB), specular microscopy (SM), and ultrasound pachymetry (UP).
Methods:
A single observer measured CCT twice in 150 eyes of 150 subjects with each of five devices: Nidek RS-3000 Advance OCT, CSO Sirius combined Scheimpflug-Placido disc system CT, Nidek AL-Scan partial coherence interferometry-based OB, Tomey EM-3000 SM, and Reichert iPac ultrasonic pachymeter. Pachymetry values corrected by the SM device software were also recorded. Levels of agreement between devices were evaluated by Bland-Altman plots with 95% limits of agreement, and repeatability for each device was analysed with intraclass correlation coefficients.
Results:
The mean CCTs measured by OCT, CT, OB, SM, corrected SM, and UP were 544.60 ± 29.56, 536.19 ± 32.14, 528.29 ± 29.45, 524.88 ± 32.38, 537.88 ± 32.38, and 545.29 ± 30.75 μm, respectively. Mean CCT differed significantly between the devices (p < 0.05) apart from between OCT and UP, and between CT and corrected SM. Mean paired differences between devices ranged between 0.68 and 20.41 μm. Repeatability with all devices was excellent (> 0.99). The range of limits of agreement was the least between OCT and UP.
Conclusions:
Different CCT measurement techniques produce quite different results, so CCT evaluation and follow-up should be performed using the same device or devices with close compatibility.
Related Concept Videos
Measures of Central Tendency
The Central Dogma
The Central Dogma
RNA is the Missing Link Between DNA and Proteins
In the early 1900s, scientists discovered that DNA stores all the information needed for cellular functions and that proteins perform most of these functions. However, the mechanisms of converting genetic information into functional proteins remained unknown for many years. Initially, it was believed that a single gene is...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
What is Central Tendency?
The central tendency is the most conventionally used data characteristic. It is a...
Trait Centrality