Monocenter Investigation Micra® MRI study (MIMICRY): feasibility study of the magnetic resonance imaging
Hermann Blessberger1,2, Daniel Kiblboeck1, Christian Reiter1
1Department of Cardiology, Med Campus III, Kepler University Hospital Linz, Medical Faculty of the Johannes Kepler University Linz, Krankenhausstr. 9, Linz, Austria.
Aims:
As in vivo real-life data are still scarce, we conducted a study to assess the safety and feasibility of cardiac magnetic resonance imaging (MRI) in patients with a leadless pacemaker system.
Methods And Results:
In this prospective non-randomized interventional trial, we enrolled 15 patients with an MRI conditional Micra® leadless pacemaker system to undergo either a 1.5 T or 3.0 T cardiac MRI scan. Clinical adverse events as well as device parameters such as pacing threshold, sensing, impedance, and battery life were assessed at baseline as well as 1 and 3 months after the scan. Device parameter changes between different time points were tested for statistical significance and compared with pre-set cut-off values. Fourteen patients underwent the cardiac MRI scan according to the protocol as well as the scheduled follow-up visits. One participant was excluded from analysis, as the MRI scan was not possible because of severe claustrophobia. Other clinical events did not occur during the scan and the follow-up period. Device parameters stayed stable and changes during the observational period were statistically not significant (changes vs. baseline: pacing threshold: 0.01 ± 0.05 V, P = 0.308, 0.01 ± 0.07 V, P = 0.419, sensing: -0.15 ± 1.11 mV, P = 0.658, -0.19 ± 1.17 mV, P = 0.800, impedance: -7.86 ± 30.7 Ohm, P = 0.447, -7.86 ± 25.77 Ohm, P = 0.183, at 1 and 3 months follow-up, respectively). Parameter changes were not statistically different between patients who underwent imaging at 1.5 T (n = 7) or 3.0 T (n = 7).
Conclusion:
In our set of patients with a Micra® leadless pacemaker, cardiac magnetic resonance imaging at either 1.5 T or 3.0 T proved feasible and safe with no relevant changes in device parameters within 3 months of follow-up.
More Related Videos
07:34Functional MRI in Conjunction with a Novel MRI-compatible Hand-induced Robotic Device to Evaluate Rehabilitation of Individuals Recovering from Hand Grip Deficits
Published on: November 23, 2019
13:35Transcranial Direct Current Stimulation and Simultaneous Functional Magnetic Resonance Imaging
Published on: April 27, 2014
Related Concept Videos
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Magnetic Resonance Imaging
Nonconscious Mimicry
Imaging Studies I: Kidney, Ureter, and Bladder Studies
Case Studies