Stented valve dynamic behavior induced by polyester fiber leaflet material in transcatheter aortic valve devices
Hoda Hatoum1, Frederick Heim2, Lakshmi Prasad Dasi1
1Department of Biomedical Engineering, the Ohio State University, Columbus, OH, United States.
Objective:
This study aims at assessing the global dynamic behavior, elastic deformability, closing energy and turbulence of rigid versus deformable stented (RS vs DS) valve systems with deformable and rigid textile materials (DT vs RT) through studying the stent-valve interaction compared to a bioprosthetic material in transcatheter aortic valves (TAV).
Methods:
Three 19 mm stented textile TAV designs (RS-DT, RS-RT and DS-RT) with different stent and leaflet properties were tested and compared with a control bioprosthetic TAV (RS-DB) in a left heart simulator flow loop under physiological pressure and flow. Particle Image Velocimetry and high speed imaging were performed. Pressure gradients (PG), leakage fractions (LF), Pinwheeling indices (PI), closing energy (E) and Reynolds shear stresses (RSS) were calculated.
Results:
(a) PGs and LFs were 11.86 ± 0.51 mmHg, 11.70 ± 0.34%; 8.84 ± 0.40 mmHg, 29.80 ± 0.76%; 11.59 ± 0.12 mmHg, 14.23 ± 1.64%; and 7.05 ± 0.09 mmHg, 12.08 ± 0.45% % for RS-DB, RS-DT, RS-RT and DS-RT respectively. (b) PIs were 15.79 ± 2.34%, 4.36 ± 0.84%, 2.47 ± 0.51% and 2.03 ± 0.33% for RS-DB, RS-DT, RS-RT and DS-RT respectively. (c) E is lowest for DS-RT (0.0010 ± 0.0002 J) followed by RS-RT (0.0017 ± 0.0002 J), RS-DB (0.0023 ± 0.0004 J) and highest with RS-DT (0.0036 ± 0.0007 J). (d) At peak systole lowest RSS was obtained with RS-DT (87.82 ± 0.58 Pa) and highest with DS-RT (122.98 ± 1.87 Pa).
Conclusion:
PGs, LFs, PIs and E were improved with DS-RT compared to other textile TAVs and RS-DB. Despite achieving more RSS than the rest of TAVs, DS-RT still falls within the same range of RSS produced by the other 2 valves and control exceeding the threshold for platelet activation.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Mitral Valve Prolapse I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse III: Nursing Management
Types of Step-Growth Polymers: Polyesters
Polyesters are commonly prepared from terephthalic acid and ethylene glycol; the crude product is known as poly(ethylene terephthalate) or PET. However, polyesters are synthesized industrially by transesterification of dimethyl terephthalate with ethylene glycol at 150 °C. The two reactants and the polymer...
Classification of Skeletal Muscle Fibers
Slow-Twitch Muscle Fibers
Slow oxidative, muscle fibers appear red due to large numbers of capillaries and high levels of...