Simultaneous quantification of Gemcitabine and Irinotecan hydrochloride in rat plasma by using high performance
L Malatesta1, D Cosco2, D Paolino3
1University of Chieti - Pescara "G. d'Annunzio", Department of Pharmacy, via dei Vestini 31, 66100 Chieti, Italy.
Abstract:
In this manuscript we aimed at the simultaneous separation and quantification of Gemcitabine and Irinotecan hydrochloride (injected both as single components and in combination) from Sprague Dawley rat plasma by using a validated method obtained through the use of a High Performance Liquid Chromatography (HPLC)-diode array detector (DAD). Gemcitabine and Irinotecan hydrochloride were detected and quantified using a Zorbax Extend C-18 column (250 mm × 4.6 mm; 5 μm particle size) in gradient elution mode. The chromatographic analyses were carried out in 15 min. The analytical mode was calibrated and validated in the concentration range from 0.1 to 18 μg/mL both for Gemcitabine and Irinotecan hydrochloride. Sprague Dawley rat plasma was used to perform the analysis. 3-methylxanthine was the internal standard. The weighted-matrix matched standard curves of Gemcitabine and Irinotecan hydrochloride showed a good linearity up to 18 μg/mL. Parallelism tests were also performed to evaluate whether the over-range samples could be analyzed after dilution without affecting the analytical performance. The intra- and inter-day precision (RSD%) values of Gemcitabine and Irinotecan hydrochloride were ≤7.14% and ≤11.5%, respectively. The intra- and inter-day trueness (Bias%) values were in the range from -11.5% to 1.70% for both drugs. The analytical mode performance was further tested after collecting Sprague Dawley rat plasma following a single-dose administration of chemotherapeutics or their association. The validated HPLC-DAD method allowed the simultaneous quantification of Gemcitabine and Irinotecan hydrochloride in the rat plasma, besides the evaluation of the pharmacokinetic parameters and drug delivery.
More Related Videos
07:32Detection of 3-Nitrotyrosine in Atmospheric Environments via a High-performance Liquid Chromatography-electrochemical Detector System
Published on: January 30, 2019
10:52Radiolabeling and Quantification of Cellular Levels of Phosphoinositides by High Performance Liquid Chromatography-coupled Flow Scintillation
Published on: January 6, 2016
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Elution Process
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...