Electromagnetic interference in cardiac electronic implants caused by novel electrical appliances emitting
Sarah Driessen1, Andreas Napp2, Kristina Schmiedchen1
1Research Center for Bioelectromagnetic Interaction (femu), Institute of Occupational, Social and Environmental Medicine, University Hospital, RWTH Aachen University, Pauwelsstr. 30, Aachen, Germany.
Abstract:
Electromagnetic fields (EMF) in the intermediate frequency (IF) range are generated by many novel electrical appliances, including electric vehicles, radiofrequency identification systems, induction hobs, or energy supply systems, such as wireless charging systems. The aim of this systematic review is to evaluate whether cardiovascular implantable electronic devices (CIEDs) are susceptible to electromagnetic interference (EMI) in the IF range (1 kHz-1 MHz). Additionally, we discuss the advantages and disadvantages of the different types of studies used to investigate EMI. Using the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-Analyses) statement, we collected and evaluated studies examining EMI in in vivo studies, in vitro studies (phantom studies, benchmark tests), and simulation studies. Our analysis revealed that cardiac implants are susceptible to malfunction induced by EMF in the IF range. Electromagnetic interference may in particular be provoked by security systems and induction hobs. The results of the studies evaluated in this systematic review further indicate that the likelihood for EMI is dependent on exposure-related parameters (field strength, frequency, and modulation) and on implant- as well as on lead-related parameters (model, type of implant, implant sensitivity setting, lead configuration, and implantation site). The review shows that the factors influencing EMI are not sufficiently characterized and EMF limit values for CIED patients cannot be derived yet. Future studies should therefore, consider exposure-related parameters as well as implant- and lead-related parameters systematically. Additionally, worst-case scenarios should be considered in all study types where possible.
More Related Videos
06:03Use of Electromagnetic Navigational Transthoracic Needle Aspiration E-TTNA for Sampling of Lung Nodules
Published on: May 23, 2015
05:11High-precision Electromagnetic Flowmeter with Empty Pipe Detection via Complex Programmable Logic Device-based Waveform Recognition
Published on: June 27, 2025
Related Concept Videos
The Electromagnetic Spectrum
The Electromagnetic Spectrum
Electromagnetic Fields
However, the observation of...
Electromagnetic Waves
Plane Electromagnetic Waves I
The EM field is assumed to be a...
Plane Electromagnetic Waves II