Related Experiment Video
Updated: Feb 8, 2026

Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
Published on: August 28, 2018
Efficacy of Automated Supplying Artery Tracking Software Using Multidetector-Row Computed Tomography Images for
Yukichi Tanahashi1, Hiroshi Kondo2, Masayoshi Yamamoto3
1Department of Radiology, Gifu University School of Medicine, 1-1 Yanagido, Gifu, 501-1194, Japan.
Purpose:
To assess the reliability of a prototype automated supplying artery tracking software (ASATS) using multidetector-row CT (MDCT) images in emergent TAE.
Materials And Methods:
Consecutive 53 patients underwent 57 sessions of emergent TAE during 7 months. Twenty-one cases were excluded due to a lack of CT data (n = 12) or negative angiographic findings (n = 9). Remaining 34 sessions of TAE and MDCT images in 32 patients (mean age 62.9 years; age range 37-92 years) were enrolled. ASATS was retrospectively conducted for the identification of supplying arteries which were confirmed with angiography (automated method). Manual modification was added as needed (semi-automated method). Two observers independently reviewed the MDCT images to detect supplying arteries (manual method). Detectability of supplying artery and time to analysis were compared among the automated, semi-automated, and manual methods by both observers.
Results:
A total of 64 bleeding sites were demonstrated on angiography. The detectability was 28 (43.8%) for automated method, 53 (82.8%) for semi-automated method, 55 (85.9%) for observer 1, and 58 (90.6%) for observer 2. Detectability of semi-automated method was significantly better than of automated method (P = 0.000) and comparable with manual method by both observers (P = 0.193 and 0.081). Average time to analysis was 185.4 s for automated method, 297.2 s for semi-automated method, 186.2 s for observer 1, and 243.7 s for observer 2.
Conclusion:
ASATS has a sufficient ability to identify supplying arteries of bleeding by adding manual modification as needed and can be used for emergent TAE.
Level Of Evidence:
Level 4, Case Control Study.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Arteries of Lower Limbs
Arteries and Arterioles
Arteries of the Upper Limbs
Overview of Systemic Arteries
Systemic circulation is the part of the cardiovascular system that carries oxygenated blood away from the heart to the body's tissues and returns deoxygenated blood back to the...