Related Experiment Video
Updated: Feb 8, 2026

Echo Particle Image Velocimetry
Published on: December 27, 2012
Optimized Echo Decorrelation Imaging Feedback for Bulk Ultrasound Ablation Control
Abstract:
Feasibility of controlling bulk ultrasound (US) thermal ablation using echo decorrelation imaging was investigated in ex vivo bovine liver. The first of two ablation and control procedures used a sequence of constant-intensity sonication cycles, ceased when the minimum echo decorrelation within a control region of interest (ROI) exceeded a predetermined threshold. The second procedure used a variable-intensity sonication sequence, with spatially averaged decorrelation as the stopping criterion. US exposures and echo decorrelation imaging were performed by a linear image-ablate array. Based on preliminary experiments, control ROIs and thresholds for the minimum-decorrelation and average-decorrelation criteria were specified. Controlled trials for the minimum-decorrelation and average-decorrelation criteria were compared with uncontrolled trials employing 9 or 18 cycles of matching sonication sequences. Lesion dimensions, treatment times, ablation rates, and areas under receiver operating characteristic curves were statistically compared. Successfully controlled trials using both criteria required significantly shorter treatment times than corresponding 18-cycle treatments, with better ablation prediction performance than uncontrolled 9-cycle and 18-cycle treatments. Either control approach resulted in greater ablation rate than corresponding 9-cycle or 18-cycle uncontrolled approaches. A post hoc analysis studied the effect of exchanging control criteria between the two series of controlled experiments. For either group, the average time needed to exceed the alternative decorrelation threshold approximately matched the average duration of successfully controlled experimental trials. These results indicate that either approach, using minimum-decorrelation or average-decorrelation criteria, is feasible for control of bulk US ablation. In addition, use of a variable-intensity sonication sequence for bulk US thermal ablation can result in larger ablated regions compared to constant-intensity sonication sequences.
More Related Videos
08:11Ultrasound Tissue Characterization of Human Achilles Tendon by Stability Quantification of Echo Patterns
Published on: September 5, 2025
07:38Real-time Monitoring of High Intensity Focused Ultrasound HIFU Ablation of In Vitro Canine Livers Using Harmonic Motion Imaging for Focused Ultrasound HMIFU
Published on: November 3, 2015
Related Concept Videos
Feedback control systems
Linear feedback systems are theoretical models that simplify analysis and design. These systems operate under the principle that their output is directly proportional to their input within certain ranges. For instance, an amplifier in a control system behaves linearly as long as the input signal remains within a specific range. However, most physical systems exhibit inherent nonlinearity...
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Feedback Inhibition
Feedback Loops
Effects of feedback
Feedback significantly modifies the gain of a control system. The gain of a system without feedback is altered by a factor of one plus GH, where G represents...
Bulk Modulus