Related Experiment Video
Updated: Feb 7, 2026

Reusable Single Cell for Iterative Epigenomic Analyses
Published on: February 11, 2022
Contamination of reusable electroencephalography electrodes: A multicenter study
Nancy M Albert1, James F Bena2, Charlotte Ciudad3
1Office of Nursing Research and Innovation, Cleveland Clinic, Cleveland OH.
Background:
Reusable electroencephalography cup electrodes and lead wires (rEEGs-CELWs) could be a source of microorganisms capable of causing hospital-acquired infections. The purpose of this study was to investigate for bacterial species of cleaned rEEGs-CELWs.
Methods:
This microbiologic evaluation involved 4 epilepsy monitoring units where rEEGs-CELWs were swabbed for bacteria using standard techniques. Analyses involved descriptive statistics and logistic regression (across sites).
Results:
Of 124 swabs, 31 (25.0%; range, 13.3%-43.3%) showed positive bacterial cultures, without between-site differences (P = .17). Bacteria were labeled by risk for hospital-acquired infection: no risk, potential risk (primarily in immunocompromised patients), and at risk (associated with infections and antibiotic resistance). At-risk bacteria species were Staphylococcus epidermidis (38.7%), Staphylococcus capitis subsp ureolyticus (3.2%), and Staphylococcus haemolyticus (9.6%). Potential-risk species were Micrococcus spp (22.6%), Acinetobacter lwoffii (6.5%), Staphylococcus hominis subsp hominis (6.5%), and Staphylococcus warneri (6.5%). Bacillus (9.6%) was the only no-risk species. Of 18 antibiotics tested on positive cultures, resistant bacteria were found in a median of 1 (range, 0-11) positive culture, equating to a 6.7% (range, 0%-61.1%) resistant antibiotic rate; no microorganisms were resistant to all antibiotics tested.
Conclusions:
Bacteria that were potential risk or at risk for infection were found on 22.6% of cleaned rEEGs-CELWs. Use of single-use electrodes and research on scalp infection and infection reduction interventions are warranted.
Related Concept Videos
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Standard Electrode Potentials
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Potentiometry: Membrane Electrodes
Case Studies