Validation of Optimum ROI Size for 123I-FP-CIT SPECT Imaging Using a 3D Mathematical Cylinder Phantom
Hideo Onishi1, Takayuki Sakai2, Osamu Shiromoto1
1Program in Health and Welfare, Graduate School of Comprehensive Scientific Research, Prefectural University of Hiroshima, Hiroshima, Japan.
Objectives:
The partial volume effect (PVE) of single-photon emission computed tomography (SPECT) on corpus striatum imaging is caused by the underestimation of specific binding ratio (SBR). A large ROI (region of interest) set using the Southampton method is independent of PVE for SBR. The present study aimed to determine the optimal ROI size with contrast and SBR for striatum images and validate the Southampton method using a three-dimensional mathematical cylinder (3D-MAC) phantom.
Methods:
We used ROIs sizes of 27, 36, 44, 51, 61, 68, and 76 mm for targets with diameters 40, 20, and 10 mm on reference and processed images reconstructed using the 3D-MAC phantom. Contrast values and SBR were compared with the theoretical values to obtain the optimal ROI size.
Results:
The contrast values in the ROI with diameters of 51 (target: 40 mm in diameter) and 44 (target: 20 mm in diameter) mm matched the theoretical values. However, this value did not correspond with the 10-mm-diameter target. The SBR matched the theoretical value with an ROI of > 44 mm in the 20-mm-diameter target; but, it was under- and overestimated under any other conditions.
Conclusion:
These results suggested that an ROI should be 2-4 folds larger than the target size without PVE, and that the Southampton method was remarkably accurate.
Related Concept Videos
Reliability and Validity
Mathematical Induction
Fundamental Mathematical Principles in Pharmacokinetics: Mathematical Expressions and Units
One significant application of mathematics in pharmacokinetics is the characterization of drug distribution through the volume of distribution...
Mathematical Modeling: Problem Solving
Fundamental Mathematical Principles in Pharmacokinetics: Calculus and Graphs
On the other hand, integral calculus focuses on...
Data Validation
Key parameters for method validation include: