[Reliability of an automatic monitor for blood pressure measurement]
David Cardona-Müller1, Fernando Grover-Páez1, Víctor Guzmán-Saldivar1
1Instituto de Terapéutica Experimental y Clínica, Departamento de Fisiología del Centro Universitario de Ciencias de la Salud, Universidad de Guadalajara, Guadalajara, México.
Abstract:
Background A correct blood pressure (BP) measurement is essential for the diagnosis and control of high BP.
Aim:
To evaluate the agreement and repeatability of BP measurements with the OMRON HEM-7320-LA device compared to a mercury sphygmomanometer.
Material And Methods:
A cross-sectional study comparing BP measurements made by two randomly selected trained nurses and an automatic oscillometric device. The mercurial sphygmomanometer was connected to the automated device via a "T" type connector and a dual-head stethoscope was used, allowing simultaneous measurements. The results were analyzed with one-factor analysis of variance, Bland-Altman's test, repeatability coefficient (RC), and intra-class correlation coefficient (ICC).
Results:
Forty-nine participants aged 56 ± 19 years were included. Nineteen had hypertension (38%). We did not observe a significant difference in either systolic (SBP) or diastolic blood pressure (DBP) pressure measurements between the observers and the device. The mean difference was -0.09 mmHg (95% confidence intervals (CI)-0.9 to 0.7) for SBP and -0.9 mmHg (95% CI -1.7 to -0.13) for DBP. The RC for SBP (6.2, 5.2 and 5.8 mmHg) and DBP (4.7, 4.2 y 5.2 mmHg) was similar between the observers and the device. The ICC for SBP was 0.990 (95% CI 0.983 to 0.995, p < 0.01) and 0.986 (95% CI 0.977 to 0.991, p < 0.01) for DBP.
Conclusions:
There was a high level of agreement and similar measurement repeatability in the measurements performed by the automatic device and the mercurial sphygmomanometer. No differences in BP measurements were observed.
Related Concept Videos
Measurement of Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...