Blood pressure measurement in special populations and circumstances
George S Stergiou1, Eamon Dolan2, Anastasios Kollias1
1Hypertension Center STRIDE-7, School of Medicine, Third Department of Medicine, Sotiria Hospital, National and Kapodistrian University of Athens, Greece.
Abstract:
According to the established validation protocols, a typical validation study of a blood pressure (BP) monitor includes general population adults with normal or elevated BP. It is recognized, however, that the automated (oscillometric) BP monitors may have different accuracy or uses in some special populations compared with adults in the general population. Thus, an automated BP monitor with proven accuracy in a general population of adults may not be accurate in a special population, and therefore separate validation is needed. Recognized special populations deserving separate validation are those for which there is theoretical, and also clinical evidence, that the accuracy of BP monitors in these groups differs from that in the general population. Young children, pregnant women (including those with preeclampsia), individuals with arm circumference >42 cm, and patients with atrial fibrillation are regarded as special populations. Adolescents, individuals older than 80 years, and patients with end-stage renal disease or diabetes mellitus have also been considered as possible special groups, but there is still inadequate evidence of altered accuracy of BP monitors in these subjects. Validation studies should be performed in special populations and evaluated separately after the BP-measuring device has successfully undergone a validation study in a general population (unless the test device is intended only for a special population). This article discusses issues relating to the measurement of BP and the diagnosis of hypertension in selected special populations, as well as in low-resource settings, where a simplified yet efficient evaluation strategy is necessary.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring pulse
Blood Pressure
Blood Pressure
The average BP in an adult is typically around 120/80 mmHg (millimeters of mercury). In this measurement, the numerator (120) indicates the systolic pressure, which is the pressure in the arteries during the contraction of the heart's ventricles as blood is expelled. The denominator (80) represents the...