Comparison of cross sectional optical coherence tomography images of elevated optic nerve heads across acquisition devices and scan protocols
Related Experiment Video
Updated: Feb 7, 2026

Quantification of Optic Nerve Cross Sectional Area on MRI: A Novel Protocol using Fiji Software
Published on: September 4, 2021
Megh D Patel1, Fareshta Khushzad1, Heather E Moss1,2
11Department of Ophthalmology, Stanford University, 2370 Watson Court, suite 200, Palo Alto, CA 94303 USA.
Background:
Optic nerve head measurements extracted from optical coherence tomography (OCT) show promise for monitoring clinical conditions with elevated optic nerve heads. The aim of this study is to compare reliability within and between raters and between image acquisition devices of optic nerve measurements derived from OCT scans in eyes with varying degrees of optic nerve elevation.
Methods:
Wide angle line scans and narrow angle radial scans through optic nerve heads were obtained using three spectral domain(SD) OCT devices on 5 subjects (6 swollen optic nerves, 4 normal optic nerves). Three raters independently semi-manually segmented the internal limiting membrane(ILM) and Bruch's membrane(BM) on each scan using customized software. One rater segmented each scan twice. Segmentations were qualitatively and quantitatively compared. Inter-rater, intra-rater and inter-device reliability was assessed for the optic nerve cross sectional area calculated from the ILM and BM segmentations using intraclass correlation coefficients and graphical comparison.
Results:
Line scans from all devices were qualitatively similar. Radial scans for which frame rate could not be adjusted were of lower quality. Intra-rater reliability for segmentation and optic nerve cross sectional area was better than inter-rater reliability, which was better than inter-device reliability, though all ICC exceeded 0.95. Reliability was not impacted by the degree of optic nerve elevation.
Conclusions:
SD-OCT devices acquired similar quality scans of the optic nerve head, with choice of scan protocol affecting the quality. For image derived markers, variability between devices was greater than that attributable to inter and intra-rater differences.
Related Concept Videos
Cross-Sectional Research
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Finding Volume Using Cross-Sectional Area
Profile Leveling and Cross Sections
Spinal Cord: Cross-sectional Anatomy
Gray Matter and its Components
Central to the gray matter is...
Properties of Enantiomers and Optical Activity

