Development, commissioning, and evaluation of a new intensity modulated minibeam proton therapy system
J B Farr1, V Moskvin1, R C Lukose1
1Department of Radiation Oncology, St. Jude Children's Research Hospital, 262 Danny Thomas Place, Memphis, TN, 38105-2794, USA.
Purpose:
To invent, design, construct, and commission an intensity modulated minibeam proton therapy system (IMMPT) without the need for physical collimation and to compare its resulting conformity to a conventional IMPT system.
Methods:
A proton therapy system (Hitachi, Ltd, Hitachi City, Japan; Model: Probeat-V) was specially modified to produce scanned minibeams without collimation. We performed integral depth dose acquisitions and calibrations using a large diameter parallel-plate ionization chamber in a scanning water phantom (PTW, Freiburg, Germany; Models: Bragg Peak ionization chamber, MP3-P). Spot size and shape was measured using radiochromic film (Ashland Advanced Materials, Bridgewater NJ; Type: EBT3), and a synthetic diamond diode type scanned point by point in air (PTW Models: MicroDiamond, MP3-P). The measured data were used as inputs to generate a Monte Carlo-based model for a commercial radiotherapy planning system (TPS) (Varian Medical Systems, Inc., Palo Alto, CA; Model: Eclipse v13.7.15). The regular ProBeat-V system (sigma ~2.5 mm) TPS model was available for comparison. A simulated base of skull case with small and medium targets proximal to brainstem was planned for both systems and compared.
Results:
The spot sigma is determined to be 1.4 mm at 221 MeV at the isocenter and below 1 mm at proximal distances. Integral depth doses were indistinguishable from the standard spot commissioning data. The TPS fit the spot profiles closely, giving a residual error maximum of 2.5% in the spot penumbra tails (below 5% of maximum) from the commissioned energies 69.4 to 221.3 MeV. The resulting IMMPT plans were more conformal than the IMPT plans due to a sharper dose gradient (90-10%) 1.5 mm smaller for the small target, and 1.3 mm for the large target, at a representative central axial water equivalent depth of 7 cm.
Conclusions:
We developed, implemented, and tested a new IMMPT system. The initial results look promising in cases where treatments can benefit from additional dose sparing to neighboring sensitive structures.
Related Concept Videos
Sound Intensity
Sound Intensity Level
The human ear can perceive an extensive range of sound intensity, necessitating the use of the logarithmic scale to define a physical quantity—the intensity level. It is a ratio of two intensities and...
Proton (¹H) NMR: Chemical Shift
Absorption signals of all the protium nuclei...
Intensity Of Electromagnetic Waves
Gene Therapy
¹H NMR of Labile Protons: Temporal Resolution
The –OH proton in alcohols typically appears in the range of δ 2 to 5 ppm but can vary depending on the specific...