In vitro biocompatibility of orthodontic miniscrews with human gingival fibroblast and SAOS-2 osteoblast cultures
Hannah Finke1,2, Bernd Koos3, Helge Fischer-Brandies4
1Department of Orthodontics, University of Tübingen, Osianderstr. 2-8, 72076, Tübingen, Germany. hannahfinke@gmx.de.
Purpose:
Miniscrews are an important choice for orthodontic anchorage. Yet reports on failures do exist, and attempts have been made to elucidate the causes. Clinical outcomes may be compromised not only by the mechanical implications of miniscrew design and the location of anchorage but also by poor biocompatibility. Hence, this study deals with the surface roughness and elemental composition of miniscrews and how these properties may affect the in vitro biocompatibility of four commercially available miniscrews.
Methods:
Most of the currently available miniscrews are made of TiAl6V4, an alloy widely considered to be biocompatible. The samples tested in this study included four similarly dimensioned TiAl6V4 products from different manufacturers: tomas® by Dentaurum, OrthoEasy® by Forestadent®, Dual Top™ by Jeil Medical/Promedia, and LOMAS by Mondeal®. The surface properties of these products were characterized by scanning electron microscopy (SEM) and energy-dispersive X‑ray spectroscopy (EDX). Cytotoxicity was assessed by 3-(4,5-dimethylthiazol-2-yl)-2,5-diphenyltetrazolium bromide (MTT) and agar overlay assays according to ISO 10993-5.
Results:
The miniscrew products were found to show variations in surface-finish quality pertaining to topography and chemical composition, with the latter departing slightly from the manufacturers' specifications. MTT assays yielded rates of cell culture viability in excess of 90%, and agar overlay assays did not reveal decoloration beyond the specimen outlines in any of the experimental groups tested.
Conclusions:
The four miniscrew products exhibited some minor, but statistically significant, differences in microtopography, alloy composition, and biological inertness. Cytotoxicity testing revealed that all four products should be considered non-cytotoxic, thus, ruling out poor biocompatibility as a cause of miniscrew failure.
More Related Videos
05:25Author Spotlight: Comparing Alveolar and Long Bone Remodeling to Explore OTM Model Potential
Published on: July 21, 2023
14:55Analysis of Minerals Produced by hFOB 1.19 and Saos-2 Cells Using Transmission Electron Microscopy with Energy Dispersive X-ray Microanalysis
Published on: June 24, 2018
Related Concept Videos
Introduction to Fibroblasts
meta-Directing Deactivators: –NO2, –CN, –CHO, –⁠CO2R, –COR, –CO2H
2° Amines to N-Nitrosamines: Reaction with NaNO2
SN2 Reaction: Kinetics
In a chemical reaction, a relationship exists between the concentration of reactants and the rate at which the reaction proceeds. The study to measure this relationship is known as the kinetics of a chemical reaction. Kinetic studies are used to deduce the rate law of a chemical reaction, which provides information about the species involved during the transition state of the rate-determining step. Thus, kinetic studies help to derive the mechanism of a...
SN2 Reaction: Mechanism
The presence of the more electronegative halogen in the substrate creates a polarized carbon-halide bond. The halide pulls the electron cloud generating an electrophilic center at the carbon atom. Thus, the carbon atom carries a partial positive charge while the halide has a...
SN2 Reaction: Transition State
When the nucleophile approaches the electrophilic carbon with its lone pairs, the halide acts as a leaving group and moves away with the electron-pair bonded to the carbon. Dotted partial bonds represent the bonds being formed or broken...