Central Corneal Edema with Scleral-Lens Wear
Young Hyun Kim1,2,3, Bo Tan2, Meng C Lin1,2
1a Vision Science Graduate Group, University of California , Berkeley , CA , USA.
Purpose:
To evaluate the safety of scleral-lens designs, we model and clinically assess central corneal edema induced by scleral-lens wear for healthy subjects.
Materials And Methods:
Central corneal swelling during scleral-lens wear is measured using optical coherence tomography (OCT). Transport resistances are modeled for oxygen diffusion through the scleral lens and post-lens tear-film (PoLTF), and into the cornea. Oxygen deficiency in the cornea activates anaerobic metabolic reactions that induce corneal edema. Oxygen permeability, carbon-dioxide permeability, settled-lens PoLTF thickness, and scleral-lens thickness are varied in the calculations to mimic different lens fits.
Results:
Transport modeling predicts that for open eyes, increasing PoLTF thickness from 50 to 400 µm increases central corneal swelling by approximately 1-1.5% when oxygen transmissibility (Dk/L) is greater than 10 hBarrer/cm (i.e., hectoBarrer/cm). Although swelling is larger for oxygen Dk/L < 10 hBarrer/cm, PoLTF thickness has minimal impact in this range. For open eye, oxygen transmissibility of the lens plays a significant role in corneal edema, but is negligible when oxygen Dk/L is > 40 hBarrer/cm. For closed eye, central corneal swelling is greater than 5% for an oxygen Dk/L range of 0-100 hBarrer/cm with typical lens-fitting parameters. For carbon-dioxide transmissibilities increasing from 50 to 250 hBarrer/cm and with a fixed oxygen Dk/L of 25 hBarrer/cm, calculated swelling diminishes by an additional 0.5%. Comparison of model calculations to clinical-swelling data is within the error range of the clinical measurements.
Conclusions:
Oxygen/metabolite transport calculations for open-eye scleral-lens wear show that typical PoLTF thicknesses fitted by clinicians (i.e., PoLTF thicknesses < 400 µm) with modern scleral lenses (i.e., oxygen Dk/L > 25 hBarrer/cm) produce corneal swelling of less than 2% in agreement with experiment. Therefore, scleral lenses prescribed today evoke less than physiological hypoxic swelling (i.e., less than 4%) for healthy corneas during open-eye. Closed-eye wear, however, appears clinically unsafe.
Related Concept Videos
The Central Dogma
The Central Dogma
RNA is the Missing Link Between DNA and Proteins
In the early 1900s, scientists discovered that DNA stores all the information needed for cellular functions and that proteins perform most of these functions. However, the mechanisms of converting genetic information into functional proteins remained unknown for many years. Initially, it was believed that a single gene is...
Measures of Central Tendency
What is Central Tendency?
The central tendency is the most conventionally used data characteristic. It is a...
Trait Centrality
Central Limit Theorem
The sample size, n, that...