Quantification of Liver Fat Content With Unenhanced MDCT: Phantom and Clinical Correlation With MRI Proton Density
Perry J Pickhardt1, Peter M Graffy1, Scott B Reeder1
11 Departments of Radiology and Medical Physics, University of Wisconsin School of Medicine and Public Health, 600 Highland Ave, Madison, WI 53705.
Objective:
The purpose of this study was to evaluate the relation between unenhanced CT liver attenuation values and MRI-derived proton density fat fraction (PDFF) for estimation of liver fat content at CT.
Materials And Methods:
A CT-MRI phantom was constructed and imaged containing 12 vials with lipid fractions ranging from 0% to 100%. For the retrospective clinical arm, 221 patients (120 men, 101 women; mean age, 54 years) underwent both unenhanced CT and chemical shift-encoded MRI of the liver between 2007 and 2017. Among these patients, 92 had more than one 120-kV CT scan for comparison. CT attenuation and MRI PDFF were derived with coregistered ROI measurements in the right hepatic lobe. The 120-kV subgroup of CT examinations performed within 1 month of MRI PDFF examinations (n = 72) served as the primary cohort for linear correlation. The effects of different tube voltage settings, time intervals between CT and MRI, and iron overload were assessed. Linear least squares regression analysis was performed.
Results:
Phantom results showed excellent linear fit between CT attenuation and MRI PDFF (r2 = 0.986). In patients, 120-kV CT performed within 1 month of MRI PDFF exhibited strong linear correlation (r2 = 0.828) that closely matched the phantom data, yielding the following clinical CT-MRI conversion formula: MRI PDFF (%) = -0.58 × CT attenuation (HU) + 38.2. Correlation worsened for CT-to-MRI intervals longer than 1 month (r2 = 0.565), and this specific relationship did not apply as well to non-120-kV settings (r2 = 0.554). For patients with multiple scans, correlation progressively worsened over time. CT-based liver fat content was underestimated in several patients with iron overload.
Conclusion:
The linear correlation between unenhanced CT attenuation and MRI PDFF allows quantification of liver fat content by means of unenhanced CT in clinical practice. As expected, correlation worsened with increasing CT-MRI time interval, variable tube voltage settings, and iron overload.
Related Concept Videos
Fats as Energy Storage Molecules
Correlation and Causation
Correlation versus Causation
If the dependent variable increases or decreases when the independent variable increases, there is a positive or negative...
Correlations
Correlation
Two variables, for example, a and b, are said to be positively correlated if both variables move in the same direction. In other words, a positive correlation exists between two variables, a and b, if:
Stereotype Content Model
Density