Inter- and intra-observer repeatability of aortic annulus measurements on screening CT for transcatheter aortic valve
Gesine Knobloch1, Sarah Sweetman1, Carrie Bartels1
1Department of Radiology, UW - School of Medicine and Public Health, Madison, WI, USA.
Objectives:
To investigate intra- and inter-observer repeatability of aortic annulus CT measurements for transcatheter aortic valve replacement (TAVR) by readers with different levels of experience and evaluate the impact of different multi-reader paradigms to improve prosthesis sizing.
Methods:
82 TAVR screening CTAs were evaluated twice by three raters with six (R1 = radiologist), three (R2 = 3D-laboratory technician) or zero (R3 = medical student) years of experience. Results were translated into hypothetical TAVR size recommendations. Intra- and inter-observer repeatability between single readers and three different multi-reader paradigms ([A]: two readers, [B]: three readers, or [C]: two readers + an optional third reader) were evaluated.
Results:
Intra-observer variability did not differ significantly (range: 50.1-67.8mm2). However, we found significant differences in mean inter-observer variance (p = 0.001). Multi-reader paradigms led to significantly increased precision (lower variability) for scenarios [B] and [C] (p = 0.03, p < 0.05). Compared to single readers, all multi-reader strategies clearly lowered the rate of discrepant device size categorization between repeated measurements (22-26% to 5-10%).
Conclusions:
Aortic annulus CT measurements for TAVR are highly reproducible. Multi-reader strategies provide higher precision than evaluations from single readers with different levels of experience and could effectively be implemented with two readers and an optional third reader (Paradigm C) in a clinical setting.
More Related Videos
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Aortic Regurgitation I: Introduction
Aortic Regurgitation III: Medical Management
Aortic Regurgitation IV: Nursing Management
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Naturalistic Observations
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...