Multicenter Evaluation of Prosthesis Oversizing of the SAPIEN 3 Transcatheter Heart Valve. Impact on Device Failure
Costanza Pellegrini1, Won-Keun Kim2, Andreas Holzamer3
1Klinik für Herz- und Kreislauferkrankungen, Deutsches Herzzentrum München, Technical University Munich, Munich, Germany.
Introduction And Objectives:
A certain degree of prosthesis oversizing (OS) is recommended for the SAPIEN 3 to achieve device success. However, an increase in OS may increase permanent pacemaker implantation (PPI) rates. We therefore investigated the influence of OS on device failure and PPI.
Methods:
A total of 804 patients were treated with SAPIEN 3 at 3 centers. Multislice computed tomography-derived OS was calculated and analyzed both as a continuous variable and categorized in 5% increments with -4% to 0% as reference.
Results:
Device failure occurred in 8.8% of patients. Median OS was lower in patients with device failure vs those with device success (+4% vs +8%; P=.038). A nonlinear risk pattern was shown for OS with a significantly reduced device failure rate within 4% to +22% of OS. There was no case of paravalvular leakage II+ between +10% to +20% of OS. The overall PPI rate was 16.2% and the median OS was significantly larger in patients with PPI (PPI: +9% vs no PPI: +7%; P = .025), while implantation depth did not vary in patients with vs without PPI (6.9±1.7 mm vs 6.6±1.9 mm; P=.101). The risk of PPI increased with increasing OS and was highest in the 2 highest categories.
Conclusions:
An increase in OS reduces the risk for device failure but increases the risk for PPI. There was no ideal range of OS to minimize both device failure and PPI.
More Related Videos
06:04Upper-extremity Approach for Secondary Access in Transfemoral Transcatheter Aortic Valve Implantation
Published on: August 8, 2025
05:31Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Heart Failure IV: Classification and Diagnostic Evaluation
Heart Failure II: Pathophysiology
Pathophysiology of Heart Failure
Heart Failure I: Introduction
Heart Failure VI: Adjunct Therapies