Related Experiment Video
Updated: Feb 7, 2026

Planar and Three-Dimensional Printing of Conductive Inks
Published on: December 9, 2011
Superior sealing effect of a three-dimensional printed modified supraglottic airway compared with the i-gel in a
Tomohiko Kimijima1, Mitsutaka Edanaga2, Michiaki Yamakage2
1Department of Anesthesiology, Sapporo Medical University School of Medicine, S 1, W 16, Chuo-ku, Sapporo, Hokkaido, 060-8543, Japan. tomokimijima@gmail.com.
Purpose:
The aim of this study was to compare the force exerted by a three-dimensional (3D) printed modified supraglottic airway (mSGA) vs. that exerted by the i-gel on a 3D printed airway model.
Methods:
After a preliminary experiment in Thiel embalmed cadavers, we created a 3D printed mSGA and five 3D printed airway models based on computed tomography data from five female Japanese patients. We compared the force exerted by the i-gel and mSGA on the larynx of the 3D printed airway models. In addition, tidal volumes with insertion of the airway devices into the 3D printed airway model and administration of different levels of pressure-controlled ventilation (PCV) were compared.
Results:
The values below indicate mean values ± SD (p value, 95% confidence interval) for the mSGA and i-gel, respectively. The forces exerted by the cuff parts were as follows: ventral: 12.5 ± 5.4 vs. 20.7 ± 3.7 N (p = 0.0001, - 10.0 to - 6.5), proximal: 1.9 ± 1.4 vs. 1.7 ± 1.3 N (p = 0.322, - 0.26 to 0.74), and dorsal parts: 6.9 ± 2.2 vs. 12.5 ± 4.8 N (p = 0.0001, - 7.9 to - 3.4), respectively. We also found significantly higher tidal volumes with the mSGA under PCV of 10, 15, and 20 cmH2O.
Conclusions:
The method of creating the mSGA that we proposed in this study can be applied to development of novel SGAs that is anatomically more suitable for pharyngolaryngeal structure.
More Related Videos
Related Concept Videos
Two-dimensional Gel Electrophoresis
The first dimension separation uses the isoelectric focusing or IEF technique performed on immobilized pH gradient (IPG) strips that separate proteins according to their isoelectric points.
Biological samples, such...
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...
Dimensional Analysis
Conversion Factors and Dimensional Analysis
The unit...
Dimensional Analysis
In fluid mechanics, dimensional...
Dimensional Analysis
Dimensional analysis allows us to analyze and compare physical quantities on a...