Force and Contact Location Measurement Errors of the VERASENSE
Stephanie J Nicolet-Petersen1, Stephen M Howell2, Maury Hull3
1Biomedical Engineering Graduate Group, University of California, Davis, 4635 2nd Ave (Building 97), Sacramento, CA 95817.
Background:
The OrthoSensor VERASENSE knee system is a commercially available instrumented tibial insert that provides real-time intraoperative measurements of tibial contact force and contact location to guide surgeons toward improving outcomes in total knee arthroplasty (TKA). However, the device has been used contrary to the manufacturer's recommendations in several studies and lacks published accuracy data. Therefore, the primary objectives of this study were to evaluate the device's error in tibial contact force when used according to and contrary to the manufacturer's recommendations, and also to evaluate the device's error in anterior-posterior (A-P) and medial-lateral (M-L) contact locations.
Methods:
The error in tibial contact force in single compartment distributed loading was evaluated by applying known forces in ranges within and exceeding that recommended by the manufacturer, with rezeroing as recommended by the manufacturer, and without rezeroing. The error in tibial contact location in single compartment concentrated loading was evaluated by applying known forces at known locations on the articular surface.
Results:
Exceeding the maximum allowable load and not rezeroing did not adversely affect the bias (i.e. average error) (p > 0.05). The maximum absolute bias without rezeroing was 2.9 lbf. Rezeroing more than doubled the bias. The maximum root mean squared error in tibial contact location was 1.5 mm in the A-P direction.
Conclusion:
The device measures tibial contact force with comparable error well above the maximum allowable load and without rezeroing, contrary to the manufacturer's instructions.
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Fundamental Attribution Error
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Contact-dependent Signaling
Gap Junctions
In animal cells, gap junctions are formed...
Random Error
Margin of Error