Cadaver-Specific Models for Finite-Element Analysis of Iliopsoas Impingement in Dual-Mobility Hip Implants
Thomas Zumbrunn1, Rajan Patel2, Michael P Duffy2
1Department of Orthopaedic Surgery, Technology Implementation Research Center, Massachusetts General Hospital, Boston, Massachusetts; Institute for Biomechanics, ETH Zurich, Zurich, Switzerland.
Background:
Joint dislocation is a major cause of failure in total hip arthroplasty. Dual-mobility implants provide a femoral head diameter that can match the native hip size for greater stability against dislocation. However, such large heads are prone to impingement against surrounding soft tissues. To address this concern, the concept of an anatomically contoured dual-mobility implant was evaluated using cadaver-specific finite-element analysis (FEA).
Methods:
The stiffness of 10 iliopsoas tendons was measured and also 3D bone models, contact pressure, and iliopsoas tendon stress were evaluated for 2 implant designs according to a previous cadaveric experiment. The iliopsoas interaction with an anatomically contoured and conventional dual-mobility implant was analyzed throughout hip flexion.
Results:
The tensile test of cadaveric iliopsoas tendons revealed an average linear stiffness of 339.4 N/mm, which was used as an input for the FEA. Tendon-liner contact pressure and tendon von Mises stress decreased with increasing hip flexion for both implants. Average contact pressure and von Mises stresses were lower in the anatomically contoured design compared with the conventional implant across all specimens and hip flexion angles.
Conclusions:
This study was built upon a previous cadaver study showing reduced tenting of the iliopsoas tendon for an anatomically contoured design compared with a conventional dual-mobility implant. The present cadaver-specific FEA study found reduced tendon-liner contact pressure and tendon stresses with contoured dual-mobility liners. Anatomical contoured design may be a solution to avoid anterior soft-tissue impingement when using hip prostheses with large femoral heads.
Related Concept Videos
Elements and Compounds
Elements
Elements are classified as atomic or molecular based on the nature of their basic units. They are unique forms of matter with specific chemical and physical properties that cannot break down into smaller substances by ordinary chemical reactions. There...
Periodic Classification of the Elements
Key Elements for Plant Nutrition
Elements: Chemical Symbols and Isotopes
Some symbols are derived from the common English name of the element; others are abbreviations of the name in another language — Latin, Greek or German. For example, the symbol for aluminum (common name)...
Classification of Elements and Compounds
Compounds are pure substances composed of two or more elements in fixed, definite proportions. Compounds are classified as ionic or molecular (covalent) based on the bonds...
The Periodic Table and Organismal Elements