On the total system error of a robotic radiosurgery system: phantom measurements, clinical evaluation and long-term
E Pantelis1, A Moutsatsos, C Antypas
1Medical Physics Laboratory, Medical School, National and Kapodistrian University of Athens, 75 Mikras Asias, 11527 Athens, Greece. CyberKnife and TomoTherapy department, Iatropolis Clinic, 54-56 Ethnikis Antistaseos, 15231 Athens, Greece.
Abstract:
The total system error (TSE) of a CyberKnife® system was measured using two phantom-based methods and one patient-based method. The standard radiochromic film (RCF) end-to-end (E2E) test using an anthropomorphic head and neck phantom and isocentric treatment delivery was used with the 6Dskull, Fiducial and Xsight® spine (XST) tracking methods. More than 200 RCF-based E2E results covering the period from installation in 2006 until 2017 were analyzed with respect to tracking method, system hardware and software versions, secondary collimation system, and years since installation. An independent polymer gel E2E method was also applied, involving a 3D printed head phantom and multiple spherical target volumes widely distributed within the brain. Finally, the TSE was assessed by comparing the delineated target in the planning computed tomography images of a patient treated for a thalamic functional target with the radiation-induced lesion defined on the six-month follow-up magnetic resonance (MR) images. Statistical analysis of the RCF-based TSE results showed mean ± standard deviation values of 0.40 ± 0.18 mm, 0.40 ± 0.19 mm, and 0.55 ± 0.20 mm for the 6Dskull, Fiducial, and XST tracking methods, respectively. Polymer gel TSE values smaller than 0.66 mm were found for seven targets distributed within the brain, showing that the targeting accuracy of the system is sustained even for targets situated up to 80 mm away from the center of the skull. An average clinical TSE value of 0.87 ± 0.25 mm was also measured using the FSE T2 and FLAIR post-treatment MR image data. Analysis of the long-term RCF-based E2E tests showed no changes of TSE over time. This study is the first to report long-term (>10 years) analysis of TSE, TSE measurement for targets positioned at large distances from the virtual machine isocenter, or a clinical assessment of TSE for the CyberKnife system. All of these measurements demonstrate TSE consistently < 1 mm.
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Fundamental Attribution Error
Random Error
Margin of Error
Long-term Depression
Irritable Bowel Syndrome II: Clinical Features and Diagnostic Evaluation
Irritable Bowel Syndrome (IBS) is classified into subtypes based on the predominant bowel habits as determined by the Bristol Stool Form Scale (BSFS). The subtypes are: