Freehand three-dimensional ultrasound system for assessment of scoliosis
Chung-Wai James Cheung1, Guang-Quan Zhou1, Siu-Yin Law1
1Interdisciplinary Division of Biomedical Engineering, The Hong Kong Polytechnic University, Hong Kong, China.
Background/Objective:
Standing radiograph with Cobb's method is routinely used to diagnose scoliosis, a medical condition defined as a lateral spine curvature > 10° with concordant vertebral rotation. However, radiation hazard and two-dimensional (2-D) viewing of 3-D anatomy restrict the application of radiograph in scoliosis examination.
Methods:
In this study, a freehand 3-D ultrasound system was developed for the radiation-free assessment of scoliosis. Bony landmarks of the spine were manually extracted from a series of ultrasound images with their spatial information recorded to form a 3-D spine model for measuring its curvature. To validate its feasibility, in vivo measurements were conducted in 28 volunteers (age: 28.0 ± 13.0 years, 9 males and 19 females). A significant linear correlation (R2 = 0.86; p < 0.001) was found between the spine curvatures as measured by Cobb's method and the 3-D ultrasound imaging with transverse process and superior articular process as landmarks. The intra- and interobserver tests indicated that the proposed method is repeatable.
Results:
The 3-D ultrasound method using bony landmarks tended to underestimate the deformity, and a proper scaling is required. Nevertheless, this study demonstrated the feasibility of the freehand 3-D ultrasound system to assess scoliosis in the standing posture with the proposed methods and 3-D spine profile.
Conclusion:
Further studies are required to understand the variations that exist between the ultrasound and radiograph results with a larger number of volunteers, and to demonstrate its potential clinical applications for monitoring of scoliosis patients. Through further clinical trials and development, the reported 3-D ultrasound imaging system can potentially be used for scoliosis mass screening and frequent monitoring of progress and treatment outcome because of its radiation-free and easy accessibility feature.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Dimensional Analysis
Conversion Factors and Dimensional Analysis
The unit...
Dimensional Analysis
In fluid mechanics, dimensional...
Dimensional Analysis
Dimensional analysis allows us to analyze and compare physical quantities on a...
Dimensional Analysis