Quantitative Analysis of Vacuum Induction Melting by Laser-induced Breakdown Spectroscopy
Published on: June 10, 2019
[Development of a Laboratory Cement Quality Analysis Apparatus Based on Laser-Induced Breakdown Spectroscopy]
Abstract:
A laboratory cement quality analysis apparatus based on laser-induced breakdown spectroscopy (LIBS) has been developed for rapid analysis of cement composition and ratio values. In this paper, the overall structure, the optical system, the sample preparation process, as well as the spectral data analysis methods are introduced. The calibration model is established with internal standard method. A comparison as to the measurement results between LIBS and X-ray fluorescence spectrometry (XRF) has been made and being analyzed. It shows that by using the LIBS apparatus, the mean absolute error of CaO, SiO2, Al2O3, and Fe2O3 in cement raw materials is 0.46%, 0.25%, 0.13%, and 0.05%, respectively, while the mean absolute error of the ratio value such as KH, SM, and IM in cement clinker is 0.02, 0.05, and 0.04, respectively. The generated cement plasmas are verified to be in the local thermal equilibrium (LTE) condition by calculating both the plasma temperature and the electron density.
More Related Videos
10:17Laser-induced Breakdown Spectroscopy: A New Approach for Nanoparticle's Mapping and Quantification in Organ Tissue
Published on: June 18, 2014
08:53Dependence of Laser-induced Breakdown Spectroscopy Results on Pulse Energies and Timing Parameters Using Soil Simulants
Published on: September 23, 2013
Related Concept Videos
Hydration of Cement
Soundness of Cement
Golgi Apparatus
Golgi Apparatus
The Golgi apparatus is a eukaryotic organelle that has a distinctive ribbon-like appearance. It is a primary sorting and dispatch station for cargo arriving from the ER. Newly arriving vesicles enter the cis face of the Golgi, closest to the ER, and are...
Portland Cement
Fineness of Cement
Direct...