Safe pseudoelastic limit range under torsional loading with Reciproc Blue
J-H Ha1, G De-Deus2, A Versluis3
1Department of Conservative Dentistry, School of Dentistry, Kyungpook National University, Daegu, Korea.
Reciproc Blue instruments exhibit a higher distortional angle at their pseudoelastic limit compared to original Reciproc files. However, the 170° pre-set angle of dedicated endodontic motors is safe for Reciproc Blue instruments during single or time-restricted use.
Area of Science:
- Endodontics
- Materials Science
- Biomedical Engineering
Background:
- Reciproc instruments are widely used in endodontic treatments.
- Reciproc Blue instruments offer enhanced flexibility and cyclic fatigue resistance.
- Understanding the mechanical properties of endodontic files under torsional load is crucial for clinical safety.
Purpose of the Study:
- To determine the distortional angle and torsional load at the pseudoelastic limit of Reciproc Blue instruments.
- To evaluate the safety of using pre-set dedicated motors (170°) with Reciproc Blue instruments.
- To compare the torsional properties of Reciproc Blue and original Reciproc instruments.
Main Methods:
- Testing of Reciproc R25 and Reciproc Blue R25 instruments under two torsional conditions: repetitive loading and single continuous rotation until fracture.
- Determination of pseudoelastic limits from torque-rotation curves.
- Statistical analysis using two-way ANOVA and t-tests.
- Examination of instrument surfaces using field-emission scanning electron microscopy (FE-SEM).
Main Results:
- Reciproc Blue instruments demonstrated a significantly greater angle at the pseudoelastic limit than original Reciproc instruments (P < 0.05).
- Repetitive torsional loading resulted in significantly lower torsional resistance compared to single continuous rotation (P < 0.05).
- Under all tested conditions, the pseudoelastic limit remained below the 170° pre-set angle of the dedicated motor.
- FE-SEM revealed microcracks and distorted machining grooves on Reciproc Blue instruments after torsional testing.
Conclusions:
- The 170° pre-set angle of dedicated endodontic motors is safe for Reciproc Blue instruments in single or time-restricted applications.
- Reciproc Blue instruments possess superior torsional properties compared to original Reciproc files.
- While Reciproc Blue instruments show resilience, microstructural analysis indicates potential for damage under prolonged or repetitive torsional stress exceeding their limit.
Related Concept Videos
Torsion of Noncircular Members
Torsional Pendulum
As long as the rigid body's angular displacement is small, its oscillation can be modeled as a linear angular oscillation. The amplitude of the oscillation is an angle. The role of mass is played...
Bending and Torsional Moments
The reaction developed in a structural element when subjected to an external force causes the element to bend. When a structural element bends upwards, it creates compressive normal forces on the top and tensile normal forces on the bottom, resulting in a couple that determines the bending...
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
Limiting Reactant
Factors Influencing Attraction IV: Reciprocity