Fluorescence photobleaching of microplastics: A cautionary tale

Kelley D Sullivan1, Valerie Gugliada1

  • 1Department of Physics and Astronomy, Ithaca College, Ithaca, NY 14850, USA.

Insights

Fluorescence microscopy for microplastic identification is hampered by photobleaching, which rapidly reduces sample fluorescence. Recovering fluorescence after bleaching is minimal, impacting further analysis.

Area of Science:

  • Environmental Science
  • Analytical Chemistry
  • Materials Science

Background:

  • Fluorescence microscopy is crucial for identifying microplastics, especially transparent and white types.
  • Photobleaching, the loss of fluorescence under illumination, poses a significant challenge in microplastic analysis.
  • Rapid photobleaching can reduce fluorescence intensity within experimental timeframes, affecting detection and quantification.

Purpose of the Study:

  • To quantify the photobleaching rates of five common plastics under fluorescence microscopy.
  • To assess the potential for fluorescence recovery in these plastics after photobleaching.
  • To inform researchers about the limitations and potential compromises when using fluorescence microscopy for microplastic identification.

Main Methods:

  • Investigated photobleaching rates and fluorescence recovery of five common plastic types.
  • Utilized fluorescence microscopy with controlled illumination intensity.
  • Measured fluorescence intensity decay over time and assessed recovery after a 3-hour period.

Main Results:

  • Statistically significant differences in photobleaching rates were observed among the plastic types.
  • In optimal conditions, fluorescence intensity decayed by 10% in 11 seconds and 33% in 230 seconds.
  • Fluorescence recovery was minimal, with a maximum of 13% observed after 3 hours.

Conclusions:

  • Photobleaching significantly diminishes fluorescence in microplastics during microscopy, even under low illumination.
  • The lack of substantial fluorescence recovery compromises the integrity of samples for subsequent analyses.
  • Researchers must exercise caution and consider photobleaching effects when planning and executing microplastic studies using fluorescence microscopy.

Related Concept Videos

Photoluminescence: Fluorescence and Phosphorescence01:23

Photoluminescence: Fluorescence and Phosphorescence

Photoluminescence is a process where a molecule absorbs light energy and re-emits it in the form of light. This phenomenon occurs when a substance absorbs photons, promoting its electrons to higher energy level excited states, followed by a relaxation process in which the electrons return to their original ground state energy levels and emit light. Photoluminescence is widely observed in various materials, including semiconductors, and organic and inorganic compounds.
A pair of electrons in a...
3.9K
Atomic Fluorescence Spectroscopy01:29

Atomic Fluorescence Spectroscopy

Atomic fluorescence spectroscopy (AFS) is an analytical technique that involves the electronic transitions of atoms in a flame, furnace, or plasma being excited by electromagnetic (EM) radiation. When these atoms absorb energy, they become excited and subsequently release energy as they return to their original state. This emitted light, or "fluorescence," is observed at a right angle to the incident beam. Both absorption and emission processes transpire at distinct wavelengths, which...
976
Confocal Fluorescence Microscopy01:16

Confocal Fluorescence Microscopy

Confocal microscopy is an advanced microscopic technique. The prime advantage of the confocal microscope over other microscopy techniques is its ability to block the out-of-focus light from the illuminated samples using pinholes. It is widely used with fluorescence optics to obtain high-resolution, sharp contrast images. Unlike optical microscopes, confocal microscopes use a focused beam of light laser to scan the entire sample surface at different z-planes. These microscopes are, therefore,...
21.2K
Variables Affecting Phosphorescence and Fluorescence01:26

Variables Affecting Phosphorescence and Fluorescence

Fluorescence and phosphorescence are essential phenomena in fields like analytical chemistry, biological imaging, and materials science, where they detect molecular properties and visualize cellular structures. Understanding the variables that influence these luminescent behaviors is crucial for maximizing accuracy and efficiency in their applications. These variables can broadly be grouped into chemical structure, solvent properties, and external conditions, each playing a distinct role in...
1.5K
Fluorescence and Phosphorescence: Instrumentation01:25

Fluorescence and Phosphorescence: Instrumentation

Fluorometers and spectrofluorometers are two types of instruments used for measuring molecular fluorescence. These instruments differ in how they select excitation and emission wavelengths and the type of light sources they utilize. Fluorometers use absorption interference filters to choose excitation and emission wavelengths. The excitation source in a fluorometer is typically a low-pressure mercury vapor lamp that emits intense lines distributed throughout the ultraviolet and visible regions.
1.6K
Super-resolution Fluorescence Microscopy01:37

Super-resolution Fluorescence Microscopy

Super-resolution fluorescence microscopy (SRFM) provides a better resolution than conventional fluorescence microscopy by reducing the point spread function (PSF). PSF is the light intensity distribution from a point that causes it to appear blurred. Due to PSF, each fluorescing point appears bigger than its actual size, and it is the PSF interference of nearby fluorophores that causes the blurred image. Various approaches to achieving higher resolution through SRFM have recently been...
14.6K