Measuring lipoprotein-associated phospholipase A2 activity in China: Protocol comparison and recalibration
Danchen Wang1, Xiuzhi Guo1, Li'an Hou1
1Department of Clinical Laboratory, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Science, Beijing, China.
Background:
Lp-PLA2 is a novel inflammation marker in cardiovascular disease. While several manufactures have registered Lp-PLA2 activity reagents, few studies have investigated the consistency among these assays. In this study, we compared and recalibrated Lp-PLA2 activity assays.
Methods:
Serum samples from 110 patients and 140 healthy individuals were collected for method comparison and reference interval validation, respectively. Fresh human serum pools (847 and 442 U/L) were used for recalibration. Lp-PLA2 activity was analyzed using all five assays with a Beckman AU 5800 analyzer. Passing-Bablok regression equations and Bland-Altman plots were used to estimate the relationship and bias among the results. A 2.5% confidence interval (CI) and 97.5% CI were used to establish a laboratory reference interval.
Results:
Assay imprecision varied from 0.8%-2.9%, while the overall coincidence rates ranged from 75.5%-98.2%. Passing-Bablok regression shows excellent linear correlation between Evermed and Diasys (R2 = 0.999), while that between Diazyme and Evermed was poor (R2 = 0.846). The R2 and correlation coefficient r among assays were 0.846-0.999 and 0.8947-0.9993, respectively. The mean bias percentages ranged from -71.5%-1.6% and -2.0%-11.6% before and after recalibration. As Diazyme and Diasys were not comparable, the Diazyme assay was not recalibrated. The reference intervals determined for Diasys, Evermed, Hengxiao, and Zybio were 184-605, 208-704, 81-328, and 273-696 U/L, respectively.
Conclusions:
Our results indicate that recalibration increased assay agreement and also highlight the need for each laboratory to establish its own reference interval for Lp-PLA2 activity.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits
Molecular Comparison of Gases, Liquids, and Solids