Prediction of ischial ramal angle for transfemoral ischial containment sockets
Matthew Quigley1, Michael P Dillon1, Richard Gd Fernandez1
11 La Trobe University, Melbourne, VIC, Australia.
Background::
A well-fitting and comfortable ischial containment socket relies on accurately replicating the transverse plane angle of the ischium and ischial ramus angle, inside the medial socket brim. Prediction of the ischial ramus angle, may provide a way to determine the ischial ramus angle without in vivo measurement.
Objectives::
To determine the accuracy with which the ischial ramus angle could be predicted and identify which variables contributed significantly to the prediction.
Study Design::
Cross-sectional study.
Methods::
Computed tomography scans were randomly sampled from a cadaveric database (n = 200). Standard multiple regression models were developed to predict the ischial ramus angle based on pelvic measures.
Results::
The regression model explained 10.5% of the variance in ischial ramus angle (p = 0.018). The standard error of the estimate was 11.32°. While regression models by sex explained a larger proportion of the variance, the resulting accuracy was not improved.
Conclusion::
The regression models explained a small proportion of variance in ischial ramus angle. The average error associated with the prediction was too large to accurately predict the ischial ramus angle for use in clinical practice. Contrary to commonly held beliefs, there was no statistically significant difference in ischial ramus angle between sexes.
Clinical Relevance:
Prediction of ischial ramus angle does not have sufficient accuracy to be clinically useful, but descriptive data may help clinicians identify casting errors and correct these in a plaster positive, knowing that the average ischial ramus angle was 32.65°±5.59° (relative to mid-sagittal plane) and does not vary between sexes.
Related Concept Videos
Predicting Molecular Geometry
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Prediction Intervals
However, the point estimate is most likely not the exact value of the population parameter, but close to it. After calculating point estimates, we construct interval estimates, called confidence intervals or prediction intervals. This prediction interval comprises a range of values unlike the point estimate and is a better predictor of the observed sample value, y.
Angle of Twist: Problem Solving
Angle Closure Glaucoma: Treatment