[Study on the accuracy of oxygen concentration of modified oxygen treatment with Venturi and humidity system]
Qiang Wei1, Bingyu Qin, Guojun He
1Department of Central Intensive Care Unit, Henan Provincial People's Hospital, Zhengzhou 450003, Henan, China (Wei Q, Qin BY, Shi Y, Sun WT, Zhu SC, Shao HZ); Department of Intensive Care Unit, the First Affiliated Hospital of Zhejiang University, Hangzhou 310006, Zhejiang, China (He GJ); Reproductive Medicine Centre, the Third Affiliated Hospital, Zhengzhou University, Zhengzhou 450003, Henan, China (Wu YY); Department of Nursing, Henan Provincial People's Hospital, Zhengzhou 450003, Henan, China (Jing MJ). Corresponding author: Shao Huanzhang,
Objective:
To verify the accuracy of oxygen concentration (FiO2) of modified oxygen treatment with Venturi and humidity system.
Methods:
Patients just after ventilator weaning and before the removal of tracheal intubation/tracheotomy tube, who admitted to the intensive care unit (ICU) of Henan Provincial People's Hospital from May 1st to December 15th in 2017, were enrolled. All patients were given a modified oxygen treatment with Venturi and humidity system, and the oxygen flow rate (Flow) of the Venturi device and the oretical value of FiO2 were adjusted according to the patient's condition. Patients were divided into five groups based on doctor's orders: Flow 3 L/min FiO2 0.24, Flow 3 L/min FiO2 0.26, Flow 6 L/min FiO2 0.28, Flow 6 L/min FiO2 0.30, Flow 9 L/min FiO2 0.35. The value of FiO2 at the inhalation end of patients of each group was measured by TSI airflow analyzer, and the consistency between the measured value of FiO2 at the inhalation end and the FiO2 marked value of Venturi was compared and analyzed.
Results:
When the FiO2 theoretical value of Venturi were adjusted to 0.24, 0.26, 0.28, 0.30, and 0.35, the measured values of FiO2 at the inhalation end of patients were 0.38±0.05, 0.38±0.05, 0.40±0.04, 0.41±0.04, and 0.77±0.11, respectively, which were all significantly higher than the theoretical value of FiO2 (all P < 0.01). The difference between the measured value of FiO2 at the inhalation side and the FiO2 value of the Venturi annotated and the difference rate were both "V"-shaped, both of which decreased with the increase in theoretical value of FiO2 to a Flow of 9 L/min and a theoretical value of FiO2 0.35, the accuracy was the worst, with the FiO2 difference of 0.42±0.11, and the FiO2 difference rate of (121.6±36.5)%.
Conclusions:
There is a difference between the measured value and the theoretical value of FiO2 at the inhalation end of the modified Venturi oxygen therapy humidification system, which needs to be paid attention to during clinical oxygen therapy.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Transport in the Blood
Oxygenic Photosynthesis
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue,...
Oxygen Requirements and Growth Patterns
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: