Biomechanical analysis between Orthofix® external fixator and different K-wire configurations for pediatric
Wen-Chao Li1, Qing-Xu Meng2, Rui-Jiang Xu3
1Department of Pediatric Surgery, Chinese People's Liberation Army General Hospital, Beijing, 100853, China. liwenchao301@163.com.
Background:
Closed reduction and percutaneous fixation are considered as the optional treatments for displaced supracondylar humerus fractures. However, there was no published report about the biomechanical analysis in Orthofix® external fixator. In this study, we developed a model of supracondylar humerus fractures and compared the biomechanical analysis of external fixator and different K-wires configurations in order to evaluate the stability of external fixator in supracondylar humerus fractures.
Methods:
We developed an anatomic humerus model by third-generation synthetic composite, and 60 synthetic humeris were osteotomized to simulate the humeral transverse supracondylar fracture. Those fractures were reduced and fixed by external fixator or K-wires, and then biomechanical analysis was performed in extension, varus, valgus, and internal and external rotation loading. A paired-sample t test was used to evaluate the distance at the fracture site between the external fixator and K-wire configurations.
Results:
During all direction loading, there was a significant statistical difference between external fixator and K-wires (P < 0.001 for all pairwise comparisons). In extension and internal rotation loading, the external fixator and three crossed K-wires had no comparable stiffness values (P = 0.572; P = 0.795), and both were significantly greater than two crossed and lateral K-wires (P < 0.05). In external rotation loading, there was no significance between the external fixator and K-wire configurations except two lateral K-wires (P > 0.05). In valgus loading, the stability of the external fixator was less than that of three crossed K-wires (P = 0.001) but was not significantly different with those of two crossed or three lateral K-wires (P = 0.126; P = 0.564). In varus loading, the stability of the external fixator was larger than those of two and three lateral K-wires (P = 0.000; P = 007).
Conclusions:
External fixator could provide enough stability for pediatric supracondylar humerus fractures without the injury of the ulnar nerve. Besides, it could enhance the rotational stiffness of the construct in rotation loading to avoid the complication of cubitus varus.
More Related Videos
10:09Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
15:11Surgical Fixation of Sternal Fractures: Preoperative Planning and a Safe Surgical Technique Using Locked Titanium Plates and Depth Limited Drilling
Published on: January 5, 2015
Related Concept Videos
Bones of the Upper Limb: Humerus
Controller Configurations
Control-system compensation involves various configurations, most commonly series or cascade compensation, in which the controller...
Electron Configurations
The relative energies of the subshells determine the order in which atomic orbitals are filled (1s, 2s, 2p, 3s, 3p,...
Electron Configuration of Multielectron Atoms
Configurations of BJT
The common base configuration is noted for its high voltage gain, positioning it as an ideal choice for single-stage amplifier circuits, such as microphone pre-amplifiers. A notable...
Stability of Equilibrium Configuration
A stable equilibrium occurs when a system tends to return to its original position when given a small displacement, and the potential energy is at its minimum. An example of a stable equilibrium is when a cantilever beam is fixed at one end and a weight is attached to the other end. If the weight...