Effect of intraocular lens decentration on image quality tested in a custom model eye
Pablo Pérez-Merino1, Susana Marcos1
1From Visual Optics and Biophotonics Laboratory, Instituto de Óptica, Consejo Superior de Investigaciones Científicas (IO-CSIC), Madrid, Spain.
Purpose:
To evaluate the effect of intraocular lens (IOL) decentration on optical aberrations with different IOL designs.
Setting:
Visual Optics and Biophotonics Laboratory, Instituto de Óptica, Consejo Superior de Investigaciones Científicas, Madrid, Spain.
Design:
Experimental study.
Methods:
The following 18 aspheric IOLs immersed in a physical model eye were measured using a laser ray-tracing aberrometer: 10 Vivinex XY1 with spherical aberration correction of -0.18 μm (V-0.18), 4 Tecnis 1P ZCB00V with spherical aberration correction of -0.27 μm (T-0.27), and 4 Acrysof 1P SN60WF with spherical aberration correction of -0.17 μm (A-0.17). The optical aberrations were evaluated with the IOL on axis and 0.4 mm and 0.7 mm laterally decentered. The laser ray tracing-measured aberrations were compared with aberrations estimated using virtual ray tracing. Retinal image quality was evaluated from the modulation transfer function (MTF) and visual Strehl ratio.
Results:
Wave aberration measurements of the same IOL were highly repetitive (deviation <0.02 μm). Astigmatism and coma increased with decentration at a rate of 0.18 μm/mm and 0.19 μm/mm for the V-0.18, 0.20 μm/mm and 0.32 μm/mm for the T-0.27, and 0.26 μm/mm and 0.39 μm/mm for the A-0.17, respectively. The 0.7 mm lateral decentration decreased the visual Strehl ratio by 2.23 times for the V-0.18, by 2.8 times for the T-0.27, and by 3.2 times for the A-0.17. The V-0.18 showed the highest MTF and visual Strehl values for all centration values.
Conclusions:
Lateral decentration resulted in decreased retinal image quality. The V-0.18 was the most immune design to optical degradation caused by decentration.
Related Concept Videos
Testing Water Quality
Quality Control
Quality control helps track data, visualize trends, and identify variations, making it easier to detect deviations that may affect the accuracy of an analysis. One way to do this is by generating a quality control chart, which...
Quality Assurance
Quality of Water
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
In Vitro Drug Dissolution: Compendial Testing Models I