Hemodynamic monitoring by pulse contour analysis during trans-catheter aortic valve replacement: A fast and easy
Francesca Ristalli1, Salvatore Mario Romano1, Miroslava Stolcova1
1Structural Interventional Cardiology, Cardio-toraco-vascular Department, Careggi University Hospital, Florence, Italy.
Background:
Residual aortic regurgitation (AR) complicates a not negligible number of transcatheter aortic valve replacement (TAVR) procedures, and its entity is not always clear at intraprocedural angiographic and echocardiographic control. We applied a minimally invasive hemodynamic monitoring system (Pressure Recording Analytical Method, PRAM) in the setting of TAVR, with the aim of identifying parameters that may help in detection and quantification of residual AR.
Methods:
We performed hemodynamic monitoring with PRAM in 43 patients undergoing trans-femoral TAVR. Investigated parameters were systolic (Psys, mm Hg), diastolic (Pdia, mm Hg), mean (MAP, mm Hg) and dicrotic pressure (Pdic, mm Hg), cardiac output (CO, L/min), stroke volume (SV, mL), cardiac cycle efficiency (CCE, Units), dP/dtmax_rad (mm Hg/ms), MAP-Pdic (mm Hg).
Results:
Procedural success was achieved in 86% of the patients; vascular complications occurred in 3 (6.9%), death in 2 (4.7%). Twenty (46.5%) patients had at least mild residual AR. CO, SV, CCE and dP/dtmax_rad changed significantly (p < 0.001) between baseline and end of procedure in the overall population, with more evident modifications in the subgroup without residual AR. MAP-Pdic variations were statistically significant only in the subgroup without AR (p = 0.05).
Conclusions:
TAVR determined an improvement in hemodynamic parameters such as CO, SV, CCE, dP/dtmax_rad. MAP-Pdic was able to discriminate patients with significant residual AR. Hemodynamic monitoring with PRAM system during TAVR is easy and fast to obtain and may help in clinical decision-making in controversial cases.
Related Concept Videos
Therapeutic Drug Monitoring: Drug Analysis Methods
Topographic Surveying and Contours
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences