Comparison of the tissue response of selective retina therapy with or without real-time feedback-controlled dosimetry
Kim Minhee1, Young-Gun Park1, Seungbum Kang2
1Department of Ophthalmology and Visual Science, Yeouido St. Mary's Hospital, College of Medicine, The Catholic University of Korea, 10, 63-ro, Yeongdeungpo-gu, Seoul, 07345, South Korea.
Purpose:
The purpose of this study is to evaluate the retinal tissue response upon selective retina therapy (SRT) with or without real-time feedback-controlled dosimetry (RFD) in rabbits.
Methods:
Eighteen eyes of nine Chinchilla Bastard rabbits were treated by SRT with or without RFD (Q-switched Nd:YLF, wavelength 527 nm, pulse duration 1.7 μs). RFD operated by optoacoustic and reflectometric methods detects the microbubbles from retinal pigment epithelium (RPE) damage in real time. After SRT, light microscopy (LM) and scanning electron microscopy (SEM) were performed at 1 h, 1 day, 3 days, 7 days, and 1 month. The RPE-damaged area on SEM was measured by ImageJ 1 h after SRT.
Results:
Without RFD, the RPE-damaged area of 76 SRT-treated spots showed a strong positive correlation with ramping pulse energy (Pearson's correlation coefficient, 0.706; P < 0.001). With RFD, there was a weak positive correlation between the RPE-damaged area of 92 spots and ramping pulse energy (Pearson's correlation coefficient, 0.211; P = 0.044). The detection rate of RFD was 91.8% by evaluating 563 SRT spots. Histology revealed that SRT with RFD produced a selectively disrupted RPE monolayer while sparing the photoreceptor layer inner segment.
Conclusions:
RFD can be useful to titrate the pulse energy of visually undetectable SRT treatment.
Related Concept Videos
Feedback control systems
Linear feedback systems are theoretical models that simplify analysis and design. These systems operate under the principle that their output is directly proportional to their input within certain ranges. For instance, an amplifier in a control system behaves linearly as long as the input signal remains within a specific range. However, most physical systems exhibit inherent nonlinearity...
Pharmacokinetic Models: Comparison and Selection Criterion
Physiological models take a detailed approach by considering specific molecular processes. They can predict drug distribution, metabolism, and elimination changes, providing a comprehensive understanding of how drugs interact with the body.
Feedback Inhibition
Dose-Response Relationship: Selectivity and Specificity
Real Time RT-PCR
The real-time quantification of the number of amplified products is...
Feedback Loops