A novel method to measure rim deformation in UHMWPE acetabular liners
Susan Partridge1, Paige Buckley1, Greg de Boer1
1School of Mechanical Engineering, University of Leeds, LS2 9JT, UK.
Abstract:
Fluoroscopy studies of total hip replacement (THR) have shown that the femoral head and acetabular cup can separate in vivo, causing edge loading on the rim of the cup. Pre-clinical testing of THR involves ISO standard motion and loading parameters that are representative of a standard walking gait. However, a requirement for more robust testing of THR has been identified and protocols for edge loading in hip simulators have been developed. This technical note describes a method to measure rim wear and deformation on ultra-high molecular weight polyethylene acetabular liners using 2D contacting profilometry and Matlab® analysis. The method is demonstrated on liners that have been subjected to edge loading in hip simulator tests and that have been retrieved at revision surgery. A quantitative and qualitative evaluation of the rim deformation was performed with good repeatability using the method.