Echocardiographic Evaluation of Atrial Communications before Transcatheter Closure
Published on: February 8, 2022
A Novel Wire Is Effective for Echo-Guiding Percutaneous Atrial Septal Defect Closure: A Preclinical Study
Yao Liu1, Gai-Li Guo1, Feng-Wen Zhang1
1Department of Cardiovascular Surgery, National Center for Cardiovascular Disease, China & Fuwai Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing 100037, China.
Objective:
To assess the effectiveness of a novel ultrasound wire for echo-guiding percutaneous atrial septal defect (ASD) closure in a sheep model.
Methods:
After right lateral thoracotomy, ASDs were created in 20 sheep by transseptal needle puncture followed by balloon dilatation. Animals were evenly randomized into 2 groups to undergo ASD closure using echography as the only imaging tool with either COOK wire (control group) or new ultrasound wire (study group). The total procedural time, passing time (time needed for the guide wire to enter the left atrium), frequency of delivery sheath dropping into the right atrium, frequency of arrhythmias, and 1-week rate of complications were compared between the two groups.
Results:
All animals survived defect creation procedures uneventfully. ASD devices were successfully implanted in all sheep. Compared with the control group, the study group had significantly (P < 0.05) lower mean procedure time (15.36 ± 4.86 versus 25.82 ± 7.85 min), lower mean passing time (2.69 ± 0.82 versus 5.58 ± 3.34 min), lower frequency of the guide wire dropping into the right atrium (0% versus 40%), and lower frequency of atrial (4.41 ± 2.61 versus 9.60 ± 3.68) or ventricular premature contractions (0.75 ± 0.36 versus 1.34 ± 0.68), respectively, without serious complications up to one week.
Conclusion:
The novel ultrasound specialized guide wire was effective in echo-guiding percutaneous ASD closure.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
23:33The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
Related Concept Videos
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Lumber Defects
Shakes are minor fractures that run along or across the wood's annual rings, while wane is...
Magnetic Field Due to Two Straight Wires
Angle Closure Glaucoma: Treatment
Household Wiring And Electrical Safety
Magnetic Field Due To A Thin Straight Wire