Comparative Dose of Contrast-Enhanced Spectral Mammography (CESM), Digital Mammography, and Digital Breast
Jordana Phillips1, Georgeta Mihai1, Sarah Esaa Hassonjee1
11 Department of Radiology, Harvard Medical School and Beth Israel Deaconess Medical Center, 330 Brookline Ave, 4th Fl, Boston, MA 02215.
Objective:
The purpose of this study is to provide a more accurate estimation of the radiation dose of contrast-enhanced spectral mammography (CESM) relative to that of 2D digital mammography and tomosynthesis using phantom and patient data and an accepted dosimetry protocol that eliminates vendor-specific average glandular dose (AGD) estimates while including breast density.
Materials And Methods:
Patient and phantom AGD estimation was performed using two vendors (system 1 and system 2) in five imaging modes, including 2D, 3D, and CESM imaging. Patient AGD was retrospectively estimated from 45 patients who underwent mammography with all imaging modes during 2012-2016. Patient and phantom AGD were estimated using accepted European and International Atomic Energy Agency protocols for dosimetry and were compared across imaging modes using a paired t test with Bonferroni correction.
Results:
Phantom data showed that the imaging modes with the lowest to highest AGDs were system 1 2D, followed by system 2 2D and system 2 3D, which had comparable values (p = 0.6), followed by system 1 CESM, and then by system 2 2D plus 3D. One hundred eighty views in 45 patients showed that the system 1 CESM AGD was 1.8 times greater than the system 1 2D AGD (p < 0.001), 1.2 times greater than the system 2 2D AGD (p < 0.001), 1.2 times greater than the system 2 3D AGD (p < 0.001), and 0.6 times less than the system 2 2D plus 3D AGD (p < 0.001).
Conclusion:
The CESM dose for system 1 is within an acceptable range as compared with other commonly performed mammographic examinations and should not preclude its use as a diagnostic breast imaging tool.
Related Concept Videos
Social Foundations of Self IV: Self in Digital Communication
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Rational Dosage Regimen: Maintenance Dose and Loading Dose
In most cases, drugs are administered repetitively or infused continuously to maintain a steady-state concentration in the body. At a steady...
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...