Comparisons of Two Microperimeters: The Clinical Value of an Extended Stimulus Range
Tiffany Arango1, Alan R Morse2, William Seiple2,3,4,5
1Department of Psychology, Northeastern University, Boston, Massachusetts.
Significance:
Fundus-guided perimetry is a common clinical tool used to measure visual field sensitivities. Comparisons between perimeters are often made despite relative differences in hardware parameters. We directly compared two perimeters using Weber contrast, which allowed us to assess the clinical gain associated with the extended stimulus range of the macular integrity assessment (MAIA).
Purpose:
The purpose of this study was to directly compare sensitivity thresholds for two microperimeters, the MAIA and Optos optical coherence tomography/scanning laser ophthalmoscope, using Weber contrast values. We also examined the clinical utility of the extended stimulus range of the MAIA.
Methods:
Six normally sighted adults with no visual field loss and 16 adults with low vision were recruited. Thresholds were measured on the MAIA and Optos using the same threshold algorithm and test points. To compare equivalent units, decibel thresholds were converted to light increments in apostilbs and then to delta increment intensities relative to each instrument's background luminance. Repeatability was assessed for normally sighted adults by testing both instruments on 3 separate days.
Results:
For normally sighted observers, mean thresholds were similar on both instruments, and repeatability within microperimeters was high. The MAIA has a 0.3-log lower contrast range and 1.37 higher contrast range. The lower contrast values did not result in lower thresholds for the normally sighted observers on the MAIA. There was a 25% increase in the number of measurable thresholds owing to the higher contrast values in low-vision observers.
Conclusions:
The higher contrast range in the MAIA yielded only a small increase in detectable thresholds for participants with visual field loss.
Related Concept Videos
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
¹H NMR: Long-Range Coupling
In alkenes, spin information is communicated via σ–π overlap, as seen in allylic (four-bond) and homoallylic (five-bond) couplings. These coupling interactions are stronger when the σ bond is parallel to the alkene...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Variation: Normal Distribution, Range, and Standard Deviation
Angle of Twist - Elastic Range