Reducing Treatment Errors Through Point-of-Care Glucometer Configuration
Jamie L Estock1, Ivan-Thibault Pham2, Holly K Curinga3
1Center for Medical Product End-user Testing, VA [US Department of Veterans Affairs] Pittsburgh Healthcare System (VAPHS), Pittsburgh.
Background:
Blood glucose (BG) testing is the most widely performed point-of-care (POC) test in a hospital setting. Multiple adverse events reported to the Food and Drug Administration (FDA) revealed that treatment decisions may be affected by information displayed on the POC glucometer's results screen. A randomized, crossover simulation study was conducted to compare two results screen configurations for ACCU-CHEK Inform II, a POC glucometer.
Methods:
Prior to the study, a heuristic evaluation of the results screen configurations and a pilot study were conducted to select the two results screen configurations for comparison. At two multicampus medical centers, 66 nurse participants experienced two computer-based simulation scenarios that asked them to interpret glucometer readings and make treatment decisions for simulated patients with 32 mg/dL BG levels and subtle symptoms of hypoglycemia. One scenario displayed a numeric value ("32 mg/dL"), and the other displayed a range abbreviation, such as "RR LO" (out of reportable range; low). Treatment errors were recorded when the participant did not treat the hypoglycemic patient with glucose or when they administered insulin.
Results:
When ACCU-CHEK Inform II displayed an "RR LO" reading, 10.6% of participants made a treatment error, including 6.7% of participants with prior training on the meaning of an "RR LO" reading. None of the participants made a treatment error when ACCU-CHEK Inform II displayed a "32 mg/dL" reading.
Conclusion:
Displaying a numeric BG reading eliminated potentially life-threating treatment errors caused by confusing range abbreviations. Manufacturers should consider these findings during future research and development of POC glucometers.
More Related Videos
Related Concept Videos
Controller Configurations
Control-system compensation involves various configurations, most commonly series or cascade compensation, in which the controller...
Electron Configurations
The relative energies of the subshells determine the order in which atomic orbitals are filled (1s, 2s, 2p, 3s, 3p,...
Electron Configuration of Multielectron Atoms
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Configurations of BJT
The common base configuration is noted for its high voltage gain, positioning it as an ideal choice for single-stage amplifier circuits, such as microphone pre-amplifiers. A notable...
Fundamental Attribution Error