Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
High-Power and Short-Duration Ablation for Pulmonary Vein Isolation: Biophysical Characterization
Eran Leshem1, Israel Zilberman2, Cory M Tschabrunn1
1Harvard-Thorndike Electrophysiology Institute, Cardiovascular Division, Department of Medicine, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts.
Objectives:
This study sought to examine the biophysical properties of high-power and short-duration (HP-SD) radiofrequency ablation for pulmonary vein isolation.
Background:
Pulmonary vein isolation is the cornerstone of atrial fibrillation ablation. However, pulmonary vein reconnection is frequent and is often the result of catheter instability, tissue edema, and a reversible nontransmural injury. We postulated that HP-SD ablation increases lesion-to-lesion uniformity and transmurality.
Methods:
This study included 20 swine and a novel open-irrigated ablation catheter with a thermocouple system able to record temperature at the catheter-tissue interface (QDOT Micro Catheter). Step 1 compared 3 HP-SD ablation settings: 90 W/4 s, 90 W/6 s, and 70 W/8 s in a thigh muscle preparation. Ablation at 90 W/4 s was identified as the best compromise between lesion size and safety parameters, with no steam-pop or char. In step 2, a total of 174 single ablation applications were performed in the beating heart and resulted in 3 (1.7%) steam-pops, all occurring at catheter-tissue interface temperature ≥85°C. Additional 233 applications at 90 W/4 s and temperature limit of 65°C were applied without steam-pop. Step 3 compared the presence of gaps and lesion transmurality in atrial lines and pulmonary vein isolation between HP-SD (90 W/4 s, T ≤65°C) and standard (25 W/20 s) ablation.
Results:
HP-SD ablation resulted in 100% contiguous lines with all transmural lesions, whereas standard ablation had linear gaps in 25% and partial thickness lesions in 29%. Ablation with HP-SD produced wider lesions (6.02 ± 0.2 mm vs. 4.43 ± 1.0 mm; p = 0.003) at similar depth (3.58 ± 0.3 mm vs. 3.53 ± 0.6 mm; p = 0.81) and improved lesion-to-lesion uniformity with comparable safety end points.
Conclusions:
In a preclinical model, HP-SD ablation (90 W/4 s, T ≤65°C) produced an improved lesion-to-lesion uniformity, linear contiguity, and transmurality at a similar safety profile of conventional ablation.
Related Concept Videos
Power System Three-Phase Short Circuits
Power
Veins
Structure of Veins:
The structure of veins is specifically designed to assist in the low-pressure transportation of...
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...
Power and Energy
Power, defined as the time rate of expending or absorbing energy, is quantified in units called watts (W). The relation between power and energy is mathematically given as
Average Power