Hemodynamic Results and Mid-term Follow-up of 850 19 to 23 mm Perimount Magna Ease Valves
Amedeo Anselmi1, Vito Giovanni Ruggieri2, Réda Belhaj Soulami1
1Division of Thoracic and Cardiovascular Surgery, Pontchaillou University Hospital, Rennes, France.
Background:
Aortic valve replacement (AVR) in small aortic roots remains a surgical dilemma with a higher risk of patient-prosthesis mismatch (PPM). The Perimount Magna Ease aortic valve (PMEAV) represents an attractive device in such cases. We examined the early hemodynamic performance, the mid-term outcomes of the PMEAV, and the impact of PPM on outcome and functional class.
Methods:
We performed a retrospective analysis of prospectively collected in-hospital data, and a prospective single-center follow-up of 849 patients who received a 19 to 23 mm PMEAV (2008-2014). Concomitant mitral or tricuspid replacement was the exclusion criterion. Early hemodynamic features were prospectively collected; mid-term follow-up was conducted according to current guidelines.
Results:
Size of implanted prosthesis was 19 mm in 11.5% of patients, 21 mm in 36.9%, and 23 mm in 51.5%. Operative mortality was 4.5% (3.1% for isolated AVR). The rate of severe and moderate PPMs was significantly higher in the 19 mm group. Follow-up was 99.9% complete (3.7 ± 2 years). Actuarial freedom from structural valve deterioration (SVD) at 5 years was 99.1%. At stratified Kaplan-Meier's analysis, PPM and age <70 years were associated with SVD. PPM was not associated with worse functional status (New York Heart Association class) or mortality at follow-up.
Conclusion:
This series shows satisfactory clinical outcomes of the PMEAV implanted in small aortic annuli at mid-term follow-up. Although PPM may occur in smaller sizes, it has limited clinical impact, and it is not associated with mid-term mortality or worse functional class. Few SVD events are evidenced; nonetheless, limited follow-up duration and its methodology need to be considered.
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Long-term Depression
Long-term Depression
Calcium Ion Concentration Mechanism
If over...
Other Nuclides: 31P, 19F, 15N NMR
While fluorine-19 and phosphorous-31 have high natural abundances (100%) and positive gyromagnetic ratios, nitrogen-15 has a low natural abundance and a negative gyromagnetic ratio. However, nitrogen-15 is still preferred over nitrogen-14 (which has a...
Mitral Valve Prolapse I: Introduction
Long-term Potentiation