Endotracheal catheter equipped with functional cuff produces clinically relevant positive end expiratory pressure: a
Jingwen Wang1,2, Zhenbo Su2,3, Jun Oto2,4
1Department of Reproductive Medicine, Gynecology and Obstetrics, The Affiliated Hospital of Jining Medical University, Jining, China.
Abstract:
Recently, we developed a novel endotracheal catheter with functional cuff (ECFC). Using such an ECFC and a regular ICU ventilator, we were able to generate clinically relevant tidal volume in a lung model and adult human sized animal. This ECFC allows co-axial ventilation without using a jet ventilator. The aim of this study was to determine if ECFC also could generate clinically relevant positive end expiratory pressure (PEEP). The experiment was conducted on a model lung and artificial trachea. Lung model respiratory mechanics were set to simulate those of an adult human being. The tip of the distal end of ECFC 14 or 19 Fr catheter was positioned in the artificial trachea 3 cm above the carina. The proximal end of ECFC was connected to an ordinary ICU ventilator. With 14 Fr catheter at respiratory rate 10 bpm, PEEP 0, 2.9, 8.2, 12.9 cmH2O was generated at preset PEEP 0, 5, 10, 15 cmH2O respectively and tidal volume was up to 393.4 ml. With 19 Fr catheter, PEEP was 0, 2.8, 7.6, 12.3 cmH2O, at preset PEEP 0, 5, 10, 15 cmH2O respectively and the tidal volume was up to 667.3 ml. With 14 Fr catheter at respiratory rate 20 bpm, PEEP was 0, 3.9, 9.6, 14.6 cmH2O at preset PEEP 0, 5, 10, 15 cmH2O respectively and tidal volume was up to 188.8 ml. With 19 Fr catheter, PEEP was 0, 3.6, 8.9, 13 cmH2O, at preset PEEP 0, 5, 10, 15 cmH2O respectively and tidal volume was up to 345.3 ml. ECFC enables clinicians to generate not only adequate tidal volume but also clinically relevant PEEP via co-axial ventilation using an ordinary ICU ventilator.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Clinically Relevant Drug Product Specifications: Methods of Establishment
Leveling Equipment
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...